Application of Point‐of‐Care Testing for Coagulation Tests: A Comparison of Consistency and Interference With a Fully Automatic Coagulation Analysis System

ABSTRACT Background Point‐of‐care testing (POCT) of coagulation is rapid, accurate, and portable. However, because it is difficult to standardize the reference intervals due to differences in metrological traceability, normalizing the results across different testing systems is challenging. The International Organization for Standardization 22870:2016 requires POCT to be compared with large automatic analyzers for medical safety. Thus, the aim of this study was to compare the correlations, consistency, and conformity of a mechanical‐based coagulation POCT system with a fully automated coagulation analyzer, and to explore the common interfering factors for each test. Methods From September 2023 to December 2023, 551 plasma samples were collected consecutively for coagulation testing by the Laboratory Medicine Department of the First Hospital of Jilin University. Prothrombin time/international normalized ratio, thrombin time (TT), activated partial thromboplastin time (APTT), and fibrinogen concentration were determined by a Sysmex CS‐5100 coagulation analyzer to serve as a control group. The coagulation results produced by POCT served as the experimental group. Passing‐Bablok, Bland–Altman, and Fourfold tables were used for correlation, consistency, and conformity, respectively. Logistic regression was used for analysis of interference. Results The correlation coefficients r of PT, INR, TT, APTT, and fibrinogen between the two testing systems were 0.866, 0.863, 0.694, 0.904, and 0.997, respectively ( p < 0.001). The overall deviations were 1.5, 0.08, 0.2, 0.8, and 0.0, respectively ( p < 0.001). The conformity were 91.95% (137/149), 93.96% (140/149), 88.44% (130/147), 89.03% (138/155), and 100.00% (100/100), respectively. The kappa values were 0.827, 0.805, 0.570, 0.768, and 1.000, respectively ( p < 0.001). More than half of the POCT measurements were higher than the analyzer measurements. Logistic regression analysis showed that jaundice (OR = 2.903, p < 0.05) and white blood cell (WBC) count (OR = 1.012, p < 0.05) were statistically significant risk factors for outliers between the two measurement systems being compared. Conclusions We found that a mechanical‐based coagulation POCT testing system showed good correlation, consistency, and conformity with automatic analyzers. The factors of hematocrit, platelet count, hemoglobin, WBC count, and jaundice may interfere with the findings of POCT, causing higher results than those produced by large coagulation analyzers.

Authors

Institutions

Publication Details

Journal
iLABMED
Published
2025-10-04
DOI
https://doi.org/10.1002/ila2.70028
Citations
1
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
1.16

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Application of Point‐of‐Care Testing for Coagulation Tests: A Comparison of Consistency and Interference With a Fully Automatic Coagulation Analysis System

Xu Wang, Yongbing Fan, Linlin Qu, Jianguo Zhu et al.
1 citations
iLABMED
Clinical Laboratory Practices and Quality Control
1.16
article

Application of Point‐of‐Care Testing for Coagulation Tests: A Comparison of Consistency and Interference With a Fully Automatic Coagulation Analysis System

Xu Wang, Yongbing Fan, Linlin Qu, Jianguo Zhu, Wei Xu, Weimeng Cui
article en
1 citations

Abstract

ABSTRACT Background Point‐of‐care testing (POCT) of coagulation is rapid, accurate, and portable. However, because it is difficult to standardize the reference intervals due to differences in metrological traceability, normalizing the results across different testing systems is challenging. The International Organization for Standardization 22870:2016 requires POCT to be compared with large automatic analyzers for medical safety. Thus, the aim of this study was to compare the correlations, consistency, and conformity of a mechanical‐based coagulation POCT system with a fully automated coagulation analyzer, and to explore the common interfering factors for each test. Methods From September 2023 to December 2023, 551 plasma samples were collected consecutively for coagulation testing by the Laboratory Medicine Department of the First Hospital of Jilin University. Prothrombin time/international normalized ratio, thrombin time (TT), activated partial thromboplastin time (APTT), and fibrinogen concentration were determined by a Sysmex CS‐5100 coagulation analyzer to serve as a control group. The coagulation results produced by POCT served as the experimental group. Passing‐Bablok, Bland–Altman, and Fourfold tables were used for correlation, consistency, and conformity, respectively. Logistic regression was used for analysis of interference. Results The correlation coefficients r of PT, INR, TT, APTT, and fibrinogen between the two testing systems were 0.866, 0.863, 0.694, 0.904, and 0.997, respectively ( p < 0.001). The overall deviations were 1.5, 0.08, 0.2, 0.8, and 0.0, respectively ( p < 0.001). The conformity were 91.95% (137/149), 93.96% (140/149), 88.44% (130/147), 89.03% (138/155), and 100.00% (100/100), respectively. The kappa values were 0.827, 0.805, 0.570, 0.768, and 1.000, respectively ( p < 0.001). More than half of the POCT measurements were higher than the analyzer measurements. Logistic regression analysis showed that jaundice (OR = 2.903, p < 0.05) and white blood cell (WBC) count (OR = 1.012, p < 0.05) were statistically significant risk factors for outliers between the two measurement systems being compared. Conclusions We found that a mechanical‐based coagulation POCT testing system showed good correlation, consistency, and conformity with automatic analyzers. The factors of hematocrit, platelet count, hemoglobin, WBC count, and jaundice may interfere with the findings of POCT, causing higher results than those produced by large coagulation analyzers.

iLABMEDVol. 3(4)
Jilin University (CN), Changchun Institute of Biological Products (CN), First Hospital of Jilin University (CN), Changchun Institute of Technology (CN), Changchun University (CN)
People's Government of Jilin Province
Openalex Percentile: Top 17%
Clinical Laboratory Practices and Quality Control
1.16
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.