Evaluation of the Effect of Acquisition Time on Image Quality of Adrenal Carcinoma PET/CT Scans

Purpose: Adrenal adenomas are best detected and understood using a Positron Emission Tomography/Computed Tomography (PET/CT) scanner. When doing PET scans, the acquisition time of 18-Fluorodeoxyglucose (18FDG) absorption is crucial as it determines the diagnostic accuracy and quality of the images. This study aimed to investigate the effects of various acquisition periods to assess the efficacy of PET/CT in identifying adrenal adenomas (1.5 vs. 3 minutes). Materials and Methods: The research included 30 patients who were thought to have adrenal adenomas. Following the 18FDG injection, PET/CT imaging was performed on each patient using one of two distinct acquisition times: 1.5 or 3 minutes. The image quality was objectively evaluated using a 5-point Likert scale. Experienced nuclear medicine professionals used consensus reading to assess diagnostic performance for adrenal adenoma identification. Results: The preliminary findings showed that compared to the 1.5-minute acquisition technique, PET/CT imaging with a 3-minute duration after 18FDG injection produced considerably superior image quality (p < 0.05). In addition, the longer acquisition time significantly increased the chance of detecting the lesion more precisely, improving the visualisation and characterisation of adrenal adenomas. With greater sensitivity and specificity, the 3-minute acquisition methodology showed better diagnostic accuracy for adrenal adenoma identification than the 1.5-minute approach. Conclusion: The study suggests that extending the acquisition time to 3 minutes improves image quality and diagnostic performance for adrenal adenoma detection, potentially improving patient care by facilitating accurate diagnosis and treatment planning.

Authors

Institutions

Publication Details

Journal
Frontiers in biomedical technologies
Published
2025-10-04
DOI
https://doi.org/10.18502/fbt.v12i4.19812
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of the Effect of Acquisition Time on Image Quality of Adrenal Carcinoma PET/CT Scans

Hiba Al-Hameed, Hiyam Adil Altaii, Zainab Salman, Hayder Kadhim Essa et al.
Frontiers in biomedical technologies
Medical Imaging Techniques and Applications
article

Evaluation of the Effect of Acquisition Time on Image Quality of Adrenal Carcinoma PET/CT Scans

Hiba Al-Hameed, Hiyam Adil Altaii, Zainab Salman, Hayder Kadhim Essa, Mustafa Ibrahim Ahmed Aldulaimy
article en

Abstract

Purpose: Adrenal adenomas are best detected and understood using a Positron Emission Tomography/Computed Tomography (PET/CT) scanner. When doing PET scans, the acquisition time of 18-Fluorodeoxyglucose (18FDG) absorption is crucial as it determines the diagnostic accuracy and quality of the images. This study aimed to investigate the effects of various acquisition periods to assess the efficacy of PET/CT in identifying adrenal adenomas (1.5 vs. 3 minutes). Materials and Methods: The research included 30 patients who were thought to have adrenal adenomas. Following the 18FDG injection, PET/CT imaging was performed on each patient using one of two distinct acquisition times: 1.5 or 3 minutes. The image quality was objectively evaluated using a 5-point Likert scale. Experienced nuclear medicine professionals used consensus reading to assess diagnostic performance for adrenal adenoma identification. Results: The preliminary findings showed that compared to the 1.5-minute acquisition technique, PET/CT imaging with a 3-minute duration after 18FDG injection produced considerably superior image quality (p < 0.05). In addition, the longer acquisition time significantly increased the chance of detecting the lesion more precisely, improving the visualisation and characterisation of adrenal adenomas. With greater sensitivity and specificity, the 3-minute acquisition methodology showed better diagnostic accuracy for adrenal adenoma identification than the 1.5-minute approach. Conclusion: The study suggests that extending the acquisition time to 3 minutes improves image quality and diagnostic performance for adrenal adenoma detection, potentially improving patient care by facilitating accurate diagnosis and treatment planning.

Frontiers in biomedical technologies
University of Mosul (IQ), University of Baghdad (IQ), Baghdad Medical City (IQ), Ministry of Health (IQ)
Openalex Percentile: Top 59%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.