Greener colonoscopy: effect of judicious carbon dioxide use and adoption of a non-leak gas/water valve on gas emissions during colonoscopy

Healthcare is responsible for ~4.4% of global carbon dioxide (CO<sub>2</sub>) emissions and endoscopy is the third largest contributor. This study aimed to quantify CO<sub>2</sub> use in colonoscopy and assess the impact of different valves and practices on emissions and costs.CO<sub>2</sub> use was measured using a mass flow meter. The study compared CO<sub>2</sub> flow using the standard gas/water valves, which continuously release CO<sub>2</sub>, with non-leak valves, which only release CO<sub>2</sub> when depressed. It also assessed the impact of judicious use of CO<sub>2</sub>. An unpaired student t test was used to calculate statistical significance.Without a colonoscope attached, CO<sub>2</sub> flow averaged 3.24 L/min. With the standard valve, flow dropped to 2.55 L/min, and with the non-leak valve, it was negligible. CO<sub>2</sub> emissions were measured intraprocedurally during 351 colonoscopies. Using a non-leak valve and/or judicious CO<sub>2</sub> application significantly reduced emissions compared with standard practice using a standard valve. This approach could reduce local emissions by >87%. Nationally, it would lead to emissions reductions of 106.5 metric tons of CO<sub>2</sub> per annum with cost savings of >£260 000.Judicious CO<sub>2</sub> application and use of a non-leak valve significantly reduced CO<sub>2</sub> emissions and costs in colonoscopy, contributing to the UK National Health Service goal of delivering a "net zero" service. We suggest turning off CO<sub>2</sub> when not needed, adopting non-leak valves, implementing this practice in other endoscopic procedures, and encouraging all endoscope manufacturers to develop similar valves.

Authors

Institutions

Publication Details

Journal
Endoscopy
Published
2025-10-03
DOI
https://doi.org/10.1055/a-2716-4818
Primary Topic
Colorectal Cancer Screening and Detection
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Greener colonoscopy: effect of judicious carbon dioxide use and adoption of a non-leak gas/water valve on gas emissions during colonoscopy

Luke Materacki, Neasa Mc Gettigan, Roland Valori, Rebecca Anderson
Endoscopy
Colorectal Cancer Screening and Detection
article

Greener colonoscopy: effect of judicious carbon dioxide use and adoption of a non-leak gas/water valve on gas emissions during colonoscopy

Luke Materacki, Neasa Mc Gettigan, Roland Valori, Rebecca Anderson
article en

Abstract

Healthcare is responsible for ~4.4% of global carbon dioxide (CO<sub>2</sub>) emissions and endoscopy is the third largest contributor. This study aimed to quantify CO<sub>2</sub> use in colonoscopy and assess the impact of different valves and practices on emissions and costs.CO<sub>2</sub> use was measured using a mass flow meter. The study compared CO<sub>2</sub> flow using the standard gas/water valves, which continuously release CO<sub>2</sub>, with non-leak valves, which only release CO<sub>2</sub> when depressed. It also assessed the impact of judicious use of CO<sub>2</sub>. An unpaired student t test was used to calculate statistical significance.Without a colonoscope attached, CO<sub>2</sub> flow averaged 3.24 L/min. With the standard valve, flow dropped to 2.55 L/min, and with the non-leak valve, it was negligible. CO<sub>2</sub> emissions were measured intraprocedurally during 351 colonoscopies. Using a non-leak valve and/or judicious CO<sub>2</sub> application significantly reduced emissions compared with standard practice using a standard valve. This approach could reduce local emissions by >87%. Nationally, it would lead to emissions reductions of 106.5 metric tons of CO<sub>2</sub> per annum with cost savings of >£260 000.Judicious CO<sub>2</sub> application and use of a non-leak valve significantly reduced CO<sub>2</sub> emissions and costs in colonoscopy, contributing to the UK National Health Service goal of delivering a "net zero" service. We suggest turning off CO<sub>2</sub> when not needed, adopting non-leak valves, implementing this practice in other endoscopic procedures, and encouraging all endoscope manufacturers to develop similar valves.

EndoscopyVol. 58(04)
Gloucestershire Hospitals NHS Foundation Trust (GB)
Openalex Percentile: Top 62%
Colorectal Cancer Screening and Detection
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Greener colonoscopy: effect of judicious carbon dioxide use and adoption of a non-leak gas/water valve on gas emissions during colonoscopy — Luke Materacki, Neasa Mc Gettigan, et al. · Endoscopy (2025) | TGRS Research Map | TGRS