Comparison of a new virtual reality-based visual field device (oculera) with humphrey field analyzer: A single-center study

BackgroundTo examine the correlation between Humphrey visual field analyzer and an automated virtual reality perimetry test (Oculera visual field analyzer) in glaucoma patients and healthy individuals.MethodsThis prospective, single-center study was conducted on 93 eyes of 93 patients, 55 of whom were glaucoma patients (glaucoma group) and 38 healthy subjects (normal group). HFA II (24-2, Swedish Interactive Threshold Algorithm) and Oculera (24-2, Oculera Interactive) tests were applied to all participants. Mean deviation (MD), pattern standard deviation (PSD), and visual field index (VFI) values of both devices were obtained. The correlation between the two devices was evaluated by statistical analyses of MD, PSD, and VFI.ResultsThe MD values were -6.36 dB (-31.8 to 2.27) for HFA II and -5.80 dB (-29.1 to 2.2) for Oculera. The difference between MDs of Oculera and HFA II was -0.56 dB. Wilcoxon signed-rank test showed a statistically significant difference in MD values between Oculera and HFA II (<i>p</i> < 0.001). However, there was a strong correlation between MDs (r = 0.932 and <i>p</i> < 0.001). With this, the results of the Bland-Altman analysis evaluating the agreement between the two tests showed that the differences between the MD measurements were not within the acceptable error range and that the results of the devices could not be used interchangeably. A strong correlation was found between the two devices in terms of PSD and VFI values (r = 0.752, r = 0.910, respectively).ConclusionsThe current findings are promising for the future use of Oculera in the diagnosis and follow-up of glaucoma patients.

Authors

Institutions

Publication Details

Journal
European Journal of Ophthalmology
Published
2025-10-03
DOI
https://doi.org/10.1177/11206721251383750
Primary Topic
Glaucoma and retinal disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of a new virtual reality-based visual field device (oculera) with humphrey field analyzer: A single-center study

Bagım Aycin Cakir, Alper Can Yılmaz, Atilla Bayer, Hayati Yılmaz et al.
European Journal of Ophthalmology
Glaucoma and retinal disorders
article

Comparison of a new virtual reality-based visual field device (oculera) with humphrey field analyzer: A single-center study

Bagım Aycin Cakir, Alper Can Yılmaz, Atilla Bayer, Hayati Yılmaz, Mehmet Talay Köylü, Osman Melih Ceylan
article en

Abstract

BackgroundTo examine the correlation between Humphrey visual field analyzer and an automated virtual reality perimetry test (Oculera visual field analyzer) in glaucoma patients and healthy individuals.MethodsThis prospective, single-center study was conducted on 93 eyes of 93 patients, 55 of whom were glaucoma patients (glaucoma group) and 38 healthy subjects (normal group). HFA II (24-2, Swedish Interactive Threshold Algorithm) and Oculera (24-2, Oculera Interactive) tests were applied to all participants. Mean deviation (MD), pattern standard deviation (PSD), and visual field index (VFI) values of both devices were obtained. The correlation between the two devices was evaluated by statistical analyses of MD, PSD, and VFI.ResultsThe MD values were -6.36 dB (-31.8 to 2.27) for HFA II and -5.80 dB (-29.1 to 2.2) for Oculera. The difference between MDs of Oculera and HFA II was -0.56 dB. Wilcoxon signed-rank test showed a statistically significant difference in MD values between Oculera and HFA II (<i>p</i> < 0.001). However, there was a strong correlation between MDs (r = 0.932 and <i>p</i> < 0.001). With this, the results of the Bland-Altman analysis evaluating the agreement between the two tests showed that the differences between the MD measurements were not within the acceptable error range and that the results of the devices could not be used interchangeably. A strong correlation was found between the two devices in terms of PSD and VFI values (r = 0.752, r = 0.910, respectively).ConclusionsThe current findings are promising for the future use of Oculera in the diagnosis and follow-up of glaucoma patients.

European Journal of OphthalmologyVol. 36(2)
Gülhane Askerî Tıp Akademisi (TR), Sağlık Bilimleri Üniversitesi (TR), Istanbul Eye Hospital (TR), He Eye Hospital (CN), Maltepe University (TR)
Openalex Percentile: Top 59%
Glaucoma and retinal disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.