Improved Measurement Method of Human Skin Temperature Based on Human Skin-like Gradient Standard Radiation Source

Infrared thermography for human skin temperature measurement, when calibrated with standard blackbodies, suffers from errors due to the mismatch in emissivity between a blackbody and human skin. This study introduces a novel calibration method utilizing a human skin-like gradient radiation source to enhance measurement accuracy. A custom radiation source with six temperature points and skin-like emissivity was developed. Thermal imagers were calibrated using this source, and their performance was compared against traditional blackbody calibration. The proposed method reduced the calibration error to 0.04 °C, a significant improvement over the 0.15 °C error obtained with blackbody calibration. Calibration with a skin-like radiation source proves superior to the blackbody method, enabling high-accuracy (less than 0.1 °C) human skin temperature measurement for improved fever screening.

Authors

Institutions

Publication Details

Journal
Thermo
Published
2025-10-02
DOI
https://doi.org/10.3390/thermo5040038
Primary Topic
Infrared Thermography in Medicine
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Improved Measurement Method of Human Skin Temperature Based on Human Skin-like Gradient Standard Radiation Source

Guojin Feng, Tianshuo Li, Ziqi Hao, Xinhua Chen et al.
Thermo
Infrared Thermography in Medicine
article

Improved Measurement Method of Human Skin Temperature Based on Human Skin-like Gradient Standard Radiation Source

Guojin Feng, Tianshuo Li, Ziqi Hao, Xinhua Chen, Zhenyuan Zhang
article en

Abstract

Infrared thermography for human skin temperature measurement, when calibrated with standard blackbodies, suffers from errors due to the mismatch in emissivity between a blackbody and human skin. This study introduces a novel calibration method utilizing a human skin-like gradient radiation source to enhance measurement accuracy. A custom radiation source with six temperature points and skin-like emissivity was developed. Thermal imagers were calibrated using this source, and their performance was compared against traditional blackbody calibration. The proposed method reduced the calibration error to 0.04 °C, a significant improvement over the 0.15 °C error obtained with blackbody calibration. Calibration with a skin-like radiation source proves superior to the blackbody method, enabling high-accuracy (less than 0.1 °C) human skin temperature measurement for improved fever screening.

ThermoVol. 5(4)
China University of Mining and Technology (CN), National Institute of Metrology (CN), Beijing University of Civil Engineering and Architecture (CN)
Openalex Percentile: Top 59%
Infrared Thermography in Medicine
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Improved Measurement Method of Human Skin Temperature Based on Human Skin-like Gradient Standard Radiation Source — Guojin Feng, Tianshuo Li, et al. · Thermo (2025) | TGRS Research Map | TGRS