Fast Trace Detection of Chlorpyrifos Vapors Using a Handheld Ion Mobility Spectrometer Operated near Ambient Temperature

Chlorpyrifos CPF (O,O-diethyl O-(3,5,6-trichloro-2-pyridyl) phosphorothioate), known also as Chlorpyrifos-ethyl, is one of the most utilized organophosphorus pesticides worldwide. Additionally, CPF could be used as a chemical warfare agent surrogate. Although its acute toxicity is not high, it is responsible for both a large number of intoxications and chronic, delayed neurological effects. In this work, it is reported for the first time the qualitative and quantitative response produced by CPF vapors, using a pocket-held Time-of-Flight Ion Mobility Spectrometer (ToF IMS) with a non-radioactive ionization source and ammonia doping, model LCD-3.2E (Smiths Detection Ltd.), operated near ambient temperature (below 30 °C). Spectra of CPF in positive ion mode included two distinct product ion peaks; thus, identification of CPF vapors by IMS relies on these peaks-the monomer M·NH<sub>4</sub><sup>+</sup> with reduced ion mobility K<sub>0</sub> = ca. 1.76 cm<sup>2</sup> V<sup>-1</sup> s<sup>-1</sup> and the dimer M<sub>2</sub>·NH<sub>4</sub><sup>+</sup> with K<sub>0</sub> = ca. 1.47 cm<sup>2</sup> V<sup>-1</sup> s<sup>-1</sup> (where M may be assignable to CPF molecule)-and positive reactant ions (Pos RIP) have K<sub>0</sub> = ca. 2.25 cm<sup>2</sup> V<sup>-1</sup> s<sup>-1</sup>. Excellent sensitivity, with a limit of detection LOD of 0.72 ppb<sub>v</sub> (10.5 μg m<sup>-3</sup>) and a limit of quantification LOQ of 2.41 ppb<sub>v</sub> (35.1 μg m<sup>-3</sup>), has been noticed; linear response was up to 100 ppb<sub>v</sub>, while saturation occurs over ca. 1000 ppb<sub>v</sub> (14.6 mg m<sup>-3</sup>). Our results demonstrate that this method provides a robust tool for both off-site and on-site detecting and quantifying CPF vapors at trace levels, which has strong implications for either industrial hygiene or forensic investigations concerning the pesticide Chlorpyrifos, as well as for monitoring of environmental contamination by organophosphorus pesticides.

Authors

Institutions

Publication Details

Journal
Toxics
Published
2025-10-02
DOI
https://doi.org/10.3390/toxics13100843
Primary Topic
Analytical Chemistry and Sensors
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Fast Trace Detection of Chlorpyrifos Vapors Using a Handheld Ion Mobility Spectrometer Operated near Ambient Temperature

Gheorghe Roşian, Tomáš Rozsypal, Victor Bocoș-Bințințan, Adrian Pătruț et al.
Toxics
Analytical Chemistry and Sensors
article

Fast Trace Detection of Chlorpyrifos Vapors Using a Handheld Ion Mobility Spectrometer Operated near Ambient Temperature

Gheorghe Roşian, Tomáš Rozsypal, Victor Bocoș-Bințințan, Adrian Pătruț, Aurel Vasile Martiniuc, Ancuța-Maria Dodea, Alin-Gabriel Moraru, Simina Vasc, Maria-Paula Bocoș-Bințințan
article en

Abstract

Chlorpyrifos CPF (O,O-diethyl O-(3,5,6-trichloro-2-pyridyl) phosphorothioate), known also as Chlorpyrifos-ethyl, is one of the most utilized organophosphorus pesticides worldwide. Additionally, CPF could be used as a chemical warfare agent surrogate. Although its acute toxicity is not high, it is responsible for both a large number of intoxications and chronic, delayed neurological effects. In this work, it is reported for the first time the qualitative and quantitative response produced by CPF vapors, using a pocket-held Time-of-Flight Ion Mobility Spectrometer (ToF IMS) with a non-radioactive ionization source and ammonia doping, model LCD-3.2E (Smiths Detection Ltd.), operated near ambient temperature (below 30 °C). Spectra of CPF in positive ion mode included two distinct product ion peaks; thus, identification of CPF vapors by IMS relies on these peaks-the monomer M·NH<sub>4</sub><sup>+</sup> with reduced ion mobility K<sub>0</sub> = ca. 1.76 cm<sup>2</sup> V<sup>-1</sup> s<sup>-1</sup> and the dimer M<sub>2</sub>·NH<sub>4</sub><sup>+</sup> with K<sub>0</sub> = ca. 1.47 cm<sup>2</sup> V<sup>-1</sup> s<sup>-1</sup> (where M may be assignable to CPF molecule)-and positive reactant ions (Pos RIP) have K<sub>0</sub> = ca. 2.25 cm<sup>2</sup> V<sup>-1</sup> s<sup>-1</sup>. Excellent sensitivity, with a limit of detection LOD of 0.72 ppb<sub>v</sub> (10.5 μg m<sup>-3</sup>) and a limit of quantification LOQ of 2.41 ppb<sub>v</sub> (35.1 μg m<sup>-3</sup>), has been noticed; linear response was up to 100 ppb<sub>v</sub>, while saturation occurs over ca. 1000 ppb<sub>v</sub> (14.6 mg m<sup>-3</sup>). Our results demonstrate that this method provides a robust tool for both off-site and on-site detecting and quantifying CPF vapors at trace levels, which has strong implications for either industrial hygiene or forensic investigations concerning the pesticide Chlorpyrifos, as well as for monitoring of environmental contamination by organophosphorus pesticides.

ToxicsVol. 13(10)
Transelectrica (Romania) (RO), University of Agricultural Sciences and Veterinary Medicine of Cluj-Napoca (RO), Babeș-Bolyai University (RO), University of Defence (CZ)
Openalex Percentile: Top 63%
Analytical Chemistry and Sensors
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.