Characterization of the Medscint HYPERSCINT scintillation dosimetry research platform with orthovoltage photon beams and a novel quasi monoenergetic beam
<i>Objective</i>. This work aimed to evaluate the dosimetric characteristics of the new HyperScint<sup>TM</sup>Research Platform 200 system (HS-RP200), with orthovoltage beams and a novel kilovoltage converging beam, converging radiotherapy and radiosurgery (CRnR) beam.<i>Approach</i>. Several dosimetric properties of the HS-RP200 detector with several orthovoltage beams were studied in air. Relative dosimetry of the HS-RP200 detector was also studied for use on the converging x-ray beam. Energy correction factors were compared between measurement, Monte Carlo, and analytical methods.<i>Main results.</i>The signal-to-noise ratio for the detector increased as the frame exposure time increased, and repeated measurements had less uncertainty when measurements were performed without stem-effect removal (WOSER). The short-term stability study of the detector's signal showed an acceptable variance of its readings to within 1% during 1 h measurements. The detector exhibited angular independence within the measurement uncertainty of 1%. The detector exhibited excellent dose linearity response with the orthovoltage beams. During an 80 d period, repeated measurements determined that the detector possesses long-term stability within a root mean square (rms) of 2.3% and 0.6% when measurements were performed with stem effect removal and WOSER, respectively. Calculated energy correction factors implied significant fluctuations of the HS-RP200 detector with the beam quality of orthovoltage beams, ranging from 3% to 66%, depending on the divergence of the effective energy of the beam from reference quality. Monte Carlo-simulated energy correction factors agreed with measurements with a maximum difference of 5.4% when measurements were performed WOSER. Lastly, the HS-RP200 detector system exhibited good relative dosimetric characteristics with the CRnR beam. A maximum rms value of 0.019 was calculated for the normalized profiles when measurements from the HS-RP200 detector were compared to measurements from film.<i>Significance</i>. The HS-RP200 detector could be used as an alternative to other detectors with orthovoltage beams, including the CRnR beam.
Authors
- Leonard Che Fru (ORCID: https://orcid.org/0000-0001-9896-207X)
- Benjamin Insley (ORCID: https://orcid.org/0000-0002-8113-0493)
- Aviad Schori
- Reza Reiazi (ORCID: https://orcid.org/0000-0003-1323-8942)
- Dirk Alan Bartkoski (ORCID: https://orcid.org/0000-0001-5792-2207)
- Dongyoen Lee (ORCID: https://orcid.org/0000-0002-9874-5806)
- Mohammad Salehpour (ORCID: https://orcid.org/0000-0002-8090-2770)
- Ramesh Tailor
Institutions
- The University of Texas MD Anderson Cancer Center (US)
Publication Details
- Journal
- Physics in Medicine and Biology
- Published
- 2025-10-02
- DOI
- https://doi.org/10.1088/1361-6560/ae0efb
- Primary Topic
- Radiation Therapy and Dosimetry
- Type
- article
- Field-Weighted Citation Impact
- 0.00