Characterization of the Medscint HYPERSCINT scintillation dosimetry research platform with orthovoltage photon beams and a novel quasi monoenergetic beam

<i>Objective</i>. This work aimed to evaluate the dosimetric characteristics of the new HyperScint<sup>TM</sup>Research Platform 200 system (HS-RP200), with orthovoltage beams and a novel kilovoltage converging beam, converging radiotherapy and radiosurgery (CRnR) beam.<i>Approach</i>. Several dosimetric properties of the HS-RP200 detector with several orthovoltage beams were studied in air. Relative dosimetry of the HS-RP200 detector was also studied for use on the converging x-ray beam. Energy correction factors were compared between measurement, Monte Carlo, and analytical methods.<i>Main results.</i>The signal-to-noise ratio for the detector increased as the frame exposure time increased, and repeated measurements had less uncertainty when measurements were performed without stem-effect removal (WOSER). The short-term stability study of the detector's signal showed an acceptable variance of its readings to within 1% during 1 h measurements. The detector exhibited angular independence within the measurement uncertainty of 1%. The detector exhibited excellent dose linearity response with the orthovoltage beams. During an 80 d period, repeated measurements determined that the detector possesses long-term stability within a root mean square (rms) of 2.3% and 0.6% when measurements were performed with stem effect removal and WOSER, respectively. Calculated energy correction factors implied significant fluctuations of the HS-RP200 detector with the beam quality of orthovoltage beams, ranging from 3% to 66%, depending on the divergence of the effective energy of the beam from reference quality. Monte Carlo-simulated energy correction factors agreed with measurements with a maximum difference of 5.4% when measurements were performed WOSER. Lastly, the HS-RP200 detector system exhibited good relative dosimetric characteristics with the CRnR beam. A maximum rms value of 0.019 was calculated for the normalized profiles when measurements from the HS-RP200 detector were compared to measurements from film.<i>Significance</i>. The HS-RP200 detector could be used as an alternative to other detectors with orthovoltage beams, including the CRnR beam.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-10-02
DOI
https://doi.org/10.1088/1361-6560/ae0efb
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Characterization of the Medscint HYPERSCINT scintillation dosimetry research platform with orthovoltage photon beams and a novel quasi monoenergetic beam

Leonard Che Fru, Benjamin Insley, Aviad Schori, Reza Reiazi et al.
Physics in Medicine and Biology
Radiation Therapy and Dosimetry
article

Characterization of the Medscint HYPERSCINT scintillation dosimetry research platform with orthovoltage photon beams and a novel quasi monoenergetic beam

Leonard Che Fru, Benjamin Insley, Aviad Schori, Reza Reiazi, Dirk Alan Bartkoski, Dongyoen Lee, Mohammad Salehpour, Ramesh Tailor
article en

Abstract

<i>Objective</i>. This work aimed to evaluate the dosimetric characteristics of the new HyperScint<sup>TM</sup>Research Platform 200 system (HS-RP200), with orthovoltage beams and a novel kilovoltage converging beam, converging radiotherapy and radiosurgery (CRnR) beam.<i>Approach</i>. Several dosimetric properties of the HS-RP200 detector with several orthovoltage beams were studied in air. Relative dosimetry of the HS-RP200 detector was also studied for use on the converging x-ray beam. Energy correction factors were compared between measurement, Monte Carlo, and analytical methods.<i>Main results.</i>The signal-to-noise ratio for the detector increased as the frame exposure time increased, and repeated measurements had less uncertainty when measurements were performed without stem-effect removal (WOSER). The short-term stability study of the detector's signal showed an acceptable variance of its readings to within 1% during 1 h measurements. The detector exhibited angular independence within the measurement uncertainty of 1%. The detector exhibited excellent dose linearity response with the orthovoltage beams. During an 80 d period, repeated measurements determined that the detector possesses long-term stability within a root mean square (rms) of 2.3% and 0.6% when measurements were performed with stem effect removal and WOSER, respectively. Calculated energy correction factors implied significant fluctuations of the HS-RP200 detector with the beam quality of orthovoltage beams, ranging from 3% to 66%, depending on the divergence of the effective energy of the beam from reference quality. Monte Carlo-simulated energy correction factors agreed with measurements with a maximum difference of 5.4% when measurements were performed WOSER. Lastly, the HS-RP200 detector system exhibited good relative dosimetric characteristics with the CRnR beam. A maximum rms value of 0.019 was calculated for the normalized profiles when measurements from the HS-RP200 detector were compared to measurements from film.<i>Significance</i>. The HS-RP200 detector could be used as an alternative to other detectors with orthovoltage beams, including the CRnR beam.

Physics in Medicine and BiologyVol. 70(20)
The University of Texas MD Anderson Cancer Center (US)
Openalex Percentile: Top 60%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.