Rapid Screening of an Anti-interference Histamine Biosensor Based on Modulation-Driven Arrays on a Chip
This study proposes a modulated pixel array design integrated with a membrane, polyvinyl chloride (PVC), aimed at accurate determination of the histamine concentration. The system employs a high-sensitivity analog front-end coupled with a double offset cancellation comparator, achieving a wide concentration range of 0.05-100 μM, a sensitivity of 5.1 ADC units per μM, and a voltage gain of approximately 12.25 mV/μM. Upon exposure to histamine in the presence of diamine oxidase (DAO), the sensor exhibited a substantial signal change of 61.6%. A significant inhibition is observed when the interference level is substantially high, specifically at a histamine-to-putrescine/cadaverine/tyramine ratio of 1:20. The high sensitivity and specificity of the proposed pixel array make it an ideal candidate for on-site histamine detection in food and fish markets. Its portable design facilitates swift screening, enabling timely intervention to reduce histamine-related health hazards and improve the effectiveness of food safety monitoring.
Authors
- Chun‐Ting Hsieh (ORCID: https://orcid.org/0009-0009-2726-8326)
- Yun-Yu Hsieh
- Wei‐Cheng Lin (ORCID: https://orcid.org/0000-0001-9248-6850)
- Wei-Lun Yen
Institutions
- Memorial Hospital of South Bend (US)
- Chang Gung University (TW)
- Chang Gung Memorial Hospital (TW)
Publication Details
- Journal
- Journal of Agricultural and Food Chemistry
- Published
- 2025-10-02
- DOI
- https://doi.org/10.1021/acs.jafc.5c08561
- Primary Topic
- Biosensors and Analytical Detection
- Type
- article
- Field-Weighted Citation Impact
- 0.00