Tooth Geometry-dependent Sensitivity Variations in L-band Electron Paramagnetic Resonance Tooth Dosimetry: A Parametric Approach Based on Quasi-tooth Models Mimicking Human Maxillary Central Incisors

L-band in vivo electron paramagnetic resonance (EPR) tooth dosimetry is a rapid, non-invasive method for determining the ionizing radiation dose of individuals. Tooth positioning and geometry are the dominant factors influencing EPR intensity. However, the delineation of the geometric parameters and their influence on EPR intensity is still unclear. This induces limitations in the availability of dosimetry for triage and leads to the imperfect consideration of geometric variations affecting EPR intensity. This study aimed to analyze the dosimetric sensitivity variations for each geometric parameter separately. Using finite element analysis, model-based parametric analyses were performed on five geometric parameters: labial enamel thickness (τlab), enamel area (Aen), aspect ratio (RA), horizontal curvature (CH), and vertical curvature (CV). The distributions of these parameters were investigated using micro-CT images from 12 human maxillary central incisors. A quasi-tooth model was proposed to mimic the tooth geometry based on the geometric parameters. The quasi-tooth model well simulated the sensitivity variations observed in the micro-CT derived tooth model. The sensitivity showed a strong linear correlation with the three geometric parameters (τlab, CH, and CV), whereas Aen exhibited a plateau around its average value (75 mm2), and no clear trend was observed for RA. After normalization, τlab, CH, and CV were identified as the dominant contributors to sensitivity variations, whereas Aen and RA had less influence. Considering the correlation between τlab and Aen (r = 0.67, p = 0.017), previous geometric correction methods using Aen may have accounted not only for the direct influence of Aen but also for the indirect correction of τlab. These results demonstrate the principles and limitations of previous geometric correction methods and imply the need for an enhanced method that incorporates multiple parameters for L-band in vivo EPR tooth dosimetry.

Authors

Institutions

Publication Details

Journal
Health Physics
Published
2025-09-26
DOI
https://doi.org/10.1097/hp.0000000000001971
Primary Topic
Radiation Effects and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Tooth Geometry-dependent Sensitivity Variations in L-band Electron Paramagnetic Resonance Tooth Dosimetry: A Parametric Approach Based on Quasi-tooth Models Mimicking Human Maxillary Central Incisors

Sung‐Joon Ye, Chang Uk Koo, Jeonghun Oh, Kwon Choi et al.
Health Physics
Radiation Effects and Dosimetry
article

Tooth Geometry-dependent Sensitivity Variations in L-band Electron Paramagnetic Resonance Tooth Dosimetry: A Parametric Approach Based on Quasi-tooth Models Mimicking Human Maxillary Central Incisors

Sung‐Joon Ye, Chang Uk Koo, Jeonghun Oh, Kwon Choi, Jong In Park, In Jung Kim
article en

Abstract

L-band in vivo electron paramagnetic resonance (EPR) tooth dosimetry is a rapid, non-invasive method for determining the ionizing radiation dose of individuals. Tooth positioning and geometry are the dominant factors influencing EPR intensity. However, the delineation of the geometric parameters and their influence on EPR intensity is still unclear. This induces limitations in the availability of dosimetry for triage and leads to the imperfect consideration of geometric variations affecting EPR intensity. This study aimed to analyze the dosimetric sensitivity variations for each geometric parameter separately. Using finite element analysis, model-based parametric analyses were performed on five geometric parameters: labial enamel thickness (τlab), enamel area (Aen), aspect ratio (RA), horizontal curvature (CH), and vertical curvature (CV). The distributions of these parameters were investigated using micro-CT images from 12 human maxillary central incisors. A quasi-tooth model was proposed to mimic the tooth geometry based on the geometric parameters. The quasi-tooth model well simulated the sensitivity variations observed in the micro-CT derived tooth model. The sensitivity showed a strong linear correlation with the three geometric parameters (τlab, CH, and CV), whereas Aen exhibited a plateau around its average value (75 mm2), and no clear trend was observed for RA. After normalization, τlab, CH, and CV were identified as the dominant contributors to sensitivity variations, whereas Aen and RA had less influence. Considering the correlation between τlab and Aen (r = 0.67, p = 0.017), previous geometric correction methods using Aen may have accounted not only for the direct influence of Aen but also for the indirect correction of τlab. These results demonstrate the principles and limitations of previous geometric correction methods and imply the need for an enhanced method that incorporates multiple parameters for L-band in vivo EPR tooth dosimetry.

Health PhysicsVol. 129(5)
Seoul National University of Science and Technology (KR), Seoul National University (KR), Korea Research Institute of Standards and Science (KR), Seoul National University Hospital (KR), Advanced Institute of Convergence Technology (KR)
Openalex Percentile: Top 68%
Radiation Effects and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.