Anode sensitivity calibration of photomultiplier tubes used as a readout of the ELOSS detector

Abstract We present and discuss the methodology and results for an anode sensitivity calibration of Photomultiplier Tubes (PMTs) that serve as the optical readout of the Energy Loss Optical Scintillation System (ELOSS). The ELOSS detector, under development at the Facility for Rare Isotope Beams (FRIB), is a gaseous Xe scintillation-based detector designed to identify the atomic number (Z) of nuclear reaction fragments by energy-loss measurement. Variation in the anode sensitivity from tube to tube of commercially available PMTs limits ELOSS from reaching its ultimate Z resolution. A methodical approach to calibrating PMT anode sensitivities to a single target value by systematic PMT gain corrections is presented. Although the calibration process is discussed in terms of its impact on ELOSS performance, the presented method is applicable to any optical detector system composed of a large number of PMTs needing anode sensitivity calibration.

Authors

Institutions

Publication Details

Journal
Journal of Instrumentation
Published
2025-10-01
DOI
https://doi.org/10.1088/1748-0221/20/10/p10007
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Anode sensitivity calibration of photomultiplier tubes used as a readout of the ELOSS detector

J. Asciutto, I. M. Harca, M. Cortesi, S. Dziubinski
Journal of Instrumentation
Radiation Detection and Scintillator Technologies
article

Anode sensitivity calibration of photomultiplier tubes used as a readout of the ELOSS detector

J. Asciutto, I. M. Harca, M. Cortesi, S. Dziubinski
article en

Abstract

Abstract We present and discuss the methodology and results for an anode sensitivity calibration of Photomultiplier Tubes (PMTs) that serve as the optical readout of the Energy Loss Optical Scintillation System (ELOSS). The ELOSS detector, under development at the Facility for Rare Isotope Beams (FRIB), is a gaseous Xe scintillation-based detector designed to identify the atomic number (Z) of nuclear reaction fragments by energy-loss measurement. Variation in the anode sensitivity from tube to tube of commercially available PMTs limits ELOSS from reaching its ultimate Z resolution. A methodical approach to calibrating PMT anode sensitivities to a single target value by systematic PMT gain corrections is presented. Although the calibration process is discussed in terms of its impact on ELOSS performance, the presented method is applicable to any optical detector system composed of a large number of PMTs needing anode sensitivity calibration.

Journal of InstrumentationVol. 20(10)
International Isotopes (United States) (US), Michigan State University (US)
Openalex Percentile: Top 63%
Radiation Detection and Scintillator Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.