Validity of a simple spillover correction for positron emission tomography measurements in the cerebrospinal fluid region

Positron emission tomography (PET) measurements in the cerebrospinal fluid (CSF) region may be overestimated because of spillover artifacts from surrounding radioactivity. In this study, we proposed a simple spillover correction method (subtraction method) and evaluated its validity. A cylindrical phantom simulating brain ventricles was used to compare the subtraction method with the geometric transfer matrix (GTM) correction approach. And the subtraction method was applied to dynamic PET images of [<sup>18</sup>F]fluorodeoxyglucose (FDG), [<sup>18</sup>F]fluorodopa (FDOPA), and [<sup>11</sup>C]raclopride (RAC), and [<sup>15</sup>O]H<sub>2</sub>O (H<sub>2</sub>O). The effects of spillover correction on CSF measurements were assessed. Both methods effectively reduced spillover artifacts in the phantom study. In dynamic PET images, after spillover correction, time-activity curves for FDG, FDOPA, and RAC approached near-zero levels in the CSF, whereas H<sub>2</sub>O continued to show increasing activity over time. This approach effectively reduces artifacts and offers the advantages of simpler volume-of-interest settings and straightforward calculation procedures.

Authors

Institutions

Publication Details

Journal
Radiological Physics and Technology
Published
2025-10-02
DOI
https://doi.org/10.1007/s12194-025-00972-5
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validity of a simple spillover correction for positron emission tomography measurements in the cerebrospinal fluid region

Shin Hibino, Emi Hayashi, Mitsuhito Mase
Radiological Physics and Technology
Medical Imaging Techniques and Applications
article

Validity of a simple spillover correction for positron emission tomography measurements in the cerebrospinal fluid region

Shin Hibino, Emi Hayashi, Mitsuhito Mase
article en

Abstract

Positron emission tomography (PET) measurements in the cerebrospinal fluid (CSF) region may be overestimated because of spillover artifacts from surrounding radioactivity. In this study, we proposed a simple spillover correction method (subtraction method) and evaluated its validity. A cylindrical phantom simulating brain ventricles was used to compare the subtraction method with the geometric transfer matrix (GTM) correction approach. And the subtraction method was applied to dynamic PET images of [<sup>18</sup>F]fluorodeoxyglucose (FDG), [<sup>18</sup>F]fluorodopa (FDOPA), and [<sup>11</sup>C]raclopride (RAC), and [<sup>15</sup>O]H<sub>2</sub>O (H<sub>2</sub>O). The effects of spillover correction on CSF measurements were assessed. Both methods effectively reduced spillover artifacts in the phantom study. In dynamic PET images, after spillover correction, time-activity curves for FDG, FDOPA, and RAC approached near-zero levels in the CSF, whereas H<sub>2</sub>O continued to show increasing activity over time. This approach effectively reduces artifacts and offers the advantages of simpler volume-of-interest settings and straightforward calculation procedures.

Radiological Physics and TechnologyVol. 18(4)
Aichi Mizuho Junior College (JP), Nagoya City University (JP)
Openalex Percentile: Top 59%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.