The Clinical Significance of Changes in the Precorneal Tear Film in Keratoconus in the Interpretation of Tomographic Methods of Corneal Examination

Relevance . Modern methods of corneal keratotopography and tomography play a key role in the diagnosis, monitoring, and surgical treatment planning of keratoconus (KC). However, the accuracy of measurements in KC may be affected by instability of the precorneal tear film (PTF) induced by dry eye syndrome (DES). Purpose : to evaluate the clinical significance of PTF changes in KC on the interpretation of corneal tomographic data and to develop an optimized diagnostic protocol for minimizing their impact. Patients and methods. The study included 115 patients (230 eyes) with bilateral non-operated KC. A comprehensive assessment of PTF status was performed (biomicroscopy of the anterior segment using vital dyes, tear film break-up time — Norn’s test, total and basal tear secretion — Schirmer’s and Jones’ tests), followed by Scheimpflug tomography (evaluating key topographic indices, kerato- and pachymetry, as well as elevation, asphericity, and wavefront aberrations) before and after instillation of sodium hyaluronate-based artificial tears. Results . The results revealed significant PTF instability in KC patients, manifested by reduced tear film break-up time and high corneal staining intensity. After temporary PTF stabilization with artificial tears, a significant decrease was observed in anterior corneal surface indices (ISV, IVA, KI, CKI, IS, IHD), keratometry values (K1, K2, Kmean, Kmax), and aberrations (RMS total, LOA, HOA). Strong correlations were found between PTF status and topographic parameters: positive with staining ( r = 0.621–0.787) and negative with tear film break-up time ( r = –0.334 to -0.787). Conclusion . PTF impairment in KC significantly distorts tomography results, potentially leading to misdiagnosis of disease progression. Instillation of artificial tears prior to examination improves measurement accuracy by minimizing PTF instability-related artifacts. Implementing standardized PTF assessment and correction in the diagnostic algorithm for KC patients enhances data reproducibility and clinical decision-making reliability.

Authors

Institutions

Publication Details

Journal
Ophthalmology in Russia
Published
2025-10-01
DOI
https://doi.org/10.18008/1816-5095-2025-3-618-624
Primary Topic
Ocular Surface and Contact Lens
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Clinical Significance of Changes in the Precorneal Tear Film in Keratoconus in the Interpretation of Tomographic Methods of Corneal Examination

E V Sukhanova, V.V. Averich, A. Yu. Slonimsky
Ophthalmology in Russia
Ocular Surface and Contact Lens
article

The Clinical Significance of Changes in the Precorneal Tear Film in Keratoconus in the Interpretation of Tomographic Methods of Corneal Examination

E V Sukhanova, V.V. Averich, A. Yu. Slonimsky
article en

Abstract

Relevance . Modern methods of corneal keratotopography and tomography play a key role in the diagnosis, monitoring, and surgical treatment planning of keratoconus (KC). However, the accuracy of measurements in KC may be affected by instability of the precorneal tear film (PTF) induced by dry eye syndrome (DES). Purpose : to evaluate the clinical significance of PTF changes in KC on the interpretation of corneal tomographic data and to develop an optimized diagnostic protocol for minimizing their impact. Patients and methods. The study included 115 patients (230 eyes) with bilateral non-operated KC. A comprehensive assessment of PTF status was performed (biomicroscopy of the anterior segment using vital dyes, tear film break-up time — Norn’s test, total and basal tear secretion — Schirmer’s and Jones’ tests), followed by Scheimpflug tomography (evaluating key topographic indices, kerato- and pachymetry, as well as elevation, asphericity, and wavefront aberrations) before and after instillation of sodium hyaluronate-based artificial tears. Results . The results revealed significant PTF instability in KC patients, manifested by reduced tear film break-up time and high corneal staining intensity. After temporary PTF stabilization with artificial tears, a significant decrease was observed in anterior corneal surface indices (ISV, IVA, KI, CKI, IS, IHD), keratometry values (K1, K2, Kmean, Kmax), and aberrations (RMS total, LOA, HOA). Strong correlations were found between PTF status and topographic parameters: positive with staining ( r = 0.621–0.787) and negative with tear film break-up time ( r = –0.334 to -0.787). Conclusion . PTF impairment in KC significantly distorts tomography results, potentially leading to misdiagnosis of disease progression. Instillation of artificial tears prior to examination improves measurement accuracy by minimizing PTF instability-related artifacts. Implementing standardized PTF assessment and correction in the diagnostic algorithm for KC patients enhances data reproducibility and clinical decision-making reliability.

Ophthalmology in RussiaVol. 22(3)
Federal State Scientific Institution Research Institute of Eye Diseases (RU)
Openalex Percentile: Top 63%
Ocular Surface and Contact Lens
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.