Incorrectly Specified Parameters of the Calculated Calibration Template and their Impact on the Accuracy of the Radionuclide Activity Measured via Gamma-Spectrometry under the Radioactive Waste Package Certification Process

The article discusses the impact of incorrectly specified parameters describing the configuration of gammaspectrometric measurements performed for RW packages when calculating the gamma-quanta registration efficiency on the measured specific activity of radionuclides. The study shows that due to incorrectly specified parameters, the measured specific activity of radionuclides in a RW package may differ from the true value by more than 100 times. The paper proposes some interlaboratory comparison tests to ensure the accuracy and reliability of gammaspectrometric measurements performed for RW packages using the calculated calibrations.

Authors

Publication Details

Journal
Radioactive waste
Published
2025-01-01
DOI
https://doi.org/10.25283/2587-9707-2025-3-27-37
Primary Topic
Graphite, nuclear technology, radiation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Incorrectly Specified Parameters of the Calculated Calibration Template and their Impact on the Accuracy of the Radionuclide Activity Measured via Gamma-Spectrometry under the Radioactive Waste Package Certification Process

A. V. Germanov, A. P. Varlakov, M. V. Batyaeva, V. V. Plisov
Radioactive waste
Graphite, nuclear technology, radiation studies
article

Incorrectly Specified Parameters of the Calculated Calibration Template and their Impact on the Accuracy of the Radionuclide Activity Measured via Gamma-Spectrometry under the Radioactive Waste Package Certification Process

A. V. Germanov, A. P. Varlakov, M. V. Batyaeva, V. V. Plisov
article en

Abstract

The article discusses the impact of incorrectly specified parameters describing the configuration of gammaspectrometric measurements performed for RW packages when calculating the gamma-quanta registration efficiency on the measured specific activity of radionuclides. The study shows that due to incorrectly specified parameters, the measured specific activity of radionuclides in a RW package may differ from the true value by more than 100 times. The paper proposes some interlaboratory comparison tests to ensure the accuracy and reliability of gammaspectrometric measurements performed for RW packages using the calculated calibrations.

Radioactive wasteVol. 32(3)
Openalex Percentile: Top 79%
Graphite, nuclear technology, radiation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.