First Experience of Using Segmental Optical Biometry for IOL Power Calculation in the Russian Federation

Purpose . To evaluate the results of segmented optical biometry for IOL power calculation in a short-term follow-up period. Patients and methods . 68 patients (90 eyes) after implantation of various IOL models were included in the study. The mean age was 70.43 ± 9.55 years (44–89 years). Women predominated in the sample ( n = 49; 72 %). Optical biometry was performed with the Argos device (Alcon, USA) and IOLMaster 700 (Zeiss, Germany). IOL calculation was performed based on Argos data using the formulas: Barrett Universal II, Cooke K6, EVO, Hill-RBF, Hoffer QST, Kane, Pearl DGS. Phacoemulsification with IOL implantation was performed using the standard technique. The observation period was 1 month. Results . One month after the operation there was a significant increase in the distance UCVA (from 0.1 (0.04; 0.20) before the operation to 0.7 (0.188; 1.00)), distance BCVA (from 0.5 (0.3; 0.8) to 1.0 (0.9; 1.0)), and a decrease in IOP from 15 (13; 16.25) to 13 (11; 15) mmHg. The absolute error was significantly lower for the Cooke K6 formula compared with Kane (0.230 (0.130; 0.389) and 0.268 (0.143; 0.405) D, respectively, p = 0.0005) and for the Hill-RBF formula compared with Hoffer QST (0.190 (0.086; 0.399) and 0.248 (0.133; 0.478) D, respectively, p = 0.0007). The share of eyes with refraction within ±0.25 D of the predicted value was significantly higher for the Hill-RBF formula compared with Kane (62.22% and 45.56%, respectively, p = 0.002). The proportion of eyes with refraction within ±0.5, ±1.0, and ±1.5 D of the predicted value did not differ significantly between the formulas. Axial length (AL) measured with Argos, was significantly longer in “short” eyes (AL≤ 22.0 mm) and shorter in “average” and “long” eyes (AL≥ 26.0 mm) compared to IOLMaster 700. Conclusion . This work presents an analysis of the first experience of using segmented optical biometry for calculation of various types of IOLs in the Russian Federation. A favorable clinical and functional effect and a high frequency of achieving target refraction values were noted. The Hill-RBF formula demonstrates some advantage over the Kane and Hoffer QST formulas, and the Cooke K6 formula over Kane.

Authors

Institutions

Publication Details

Journal
Ophthalmology in Russia
Published
2025-10-01
DOI
https://doi.org/10.18008/1816-5095-2025-3-582-588
Citations
1
Primary Topic
Ocular and Laser Science Research
Type
article
Field-Weighted Citation Impact
1.71
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

First Experience of Using Segmental Optical Biometry for IOL Power Calculation in the Russian Federation

К. Б. Першин, Н. Ф. Пашинова, А. А. Панов, А. Yu. Tsygankov
1 citations
Ophthalmology in Russia
Ocular and Laser Science Research
1.71
article

First Experience of Using Segmental Optical Biometry for IOL Power Calculation in the Russian Federation

К. Б. Першин, Н. Ф. Пашинова, А. А. Панов, А. Yu. Tsygankov
article en
1 citations

Abstract

Purpose . To evaluate the results of segmented optical biometry for IOL power calculation in a short-term follow-up period. Patients and methods . 68 patients (90 eyes) after implantation of various IOL models were included in the study. The mean age was 70.43 ± 9.55 years (44–89 years). Women predominated in the sample ( n = 49; 72 %). Optical biometry was performed with the Argos device (Alcon, USA) and IOLMaster 700 (Zeiss, Germany). IOL calculation was performed based on Argos data using the formulas: Barrett Universal II, Cooke K6, EVO, Hill-RBF, Hoffer QST, Kane, Pearl DGS. Phacoemulsification with IOL implantation was performed using the standard technique. The observation period was 1 month. Results . One month after the operation there was a significant increase in the distance UCVA (from 0.1 (0.04; 0.20) before the operation to 0.7 (0.188; 1.00)), distance BCVA (from 0.5 (0.3; 0.8) to 1.0 (0.9; 1.0)), and a decrease in IOP from 15 (13; 16.25) to 13 (11; 15) mmHg. The absolute error was significantly lower for the Cooke K6 formula compared with Kane (0.230 (0.130; 0.389) and 0.268 (0.143; 0.405) D, respectively, p = 0.0005) and for the Hill-RBF formula compared with Hoffer QST (0.190 (0.086; 0.399) and 0.248 (0.133; 0.478) D, respectively, p = 0.0007). The share of eyes with refraction within ±0.25 D of the predicted value was significantly higher for the Hill-RBF formula compared with Kane (62.22% and 45.56%, respectively, p = 0.002). The proportion of eyes with refraction within ±0.5, ±1.0, and ±1.5 D of the predicted value did not differ significantly between the formulas. Axial length (AL) measured with Argos, was significantly longer in “short” eyes (AL≤ 22.0 mm) and shorter in “average” and “long” eyes (AL≥ 26.0 mm) compared to IOLMaster 700. Conclusion . This work presents an analysis of the first experience of using segmented optical biometry for calculation of various types of IOLs in the Russian Federation. A favorable clinical and functional effect and a high frequency of achieving target refraction values were noted. The Hill-RBF formula demonstrates some advantage over the Kane and Hoffer QST formulas, and the Cooke K6 formula over Kane.

Ophthalmology in RussiaVol. 22(3)
Eye Center (SA), Federal Medical-Biological Agency (RU)
Openalex Percentile: Top 13%
Ocular and Laser Science Research
1.71
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.