Method development for analysis pentachlorobenzene and hexachlorobutadiene in wastewater by GC/MS
This research presents the optimization of Pentachlorobenzene (PeCB) and Hexachlorobutadiene (HCBD) analysis in surface water.The study investigates the operating parameters of the GC/MS system and the clean-up, enrichment procedures based on liquid-liquid extraction using surrogate standard (Phenanthrene-D10).The calibration curve demonstrated a strong linear correlation (~0.994) within the concentration range of 25 -600g/L.The method detection limits were 0.008g/L for PeCB and 0.004g/L for HCBD.The recovery of the analytical procedure ranged from 85% to 114%.The method was applied to waste water samples showed good repeatability.
Publication Details
- Journal
- Journal of Science and Technology - HaUI
- Published
- 2025-05-31
- DOI
- https://doi.org/10.57001/huih5804.2025.178
- Primary Topic
- Analytical chemistry methods development
- Type
- article
- Field-Weighted Citation Impact
- 0.00