B-169 A Novel Immunoprecipitation-based Liquid Chromatography–Tandem Mass Spectrometry Analysis for Accurate Determination for Copeptin in Human Serum

Abstract Background Copeptin, a glycosylated peptide of 39 amino acids, is derived from the C-terminal segment of the arginine vasopressin (AVP) precursor peptide. It is released in an equimolar mode with AVP from the pituitary gland and has emerged as a stable and simple-to-measure surrogate marker for AVP. Copeptin demonstrates excellent clinical values, particularly in diagnosing polyuria-polydipsia syndromes such as diabetes insipidus. However, conventional immunoassay methods are limited and lack comparability. Therefore, our aim was to establish a liquid chromatography–tandem mass spectrometry (LC–MS/MS) method for quantifying copeptin levels in serum. Methods Copeptin was extracted from serum using immunoprecipitation with antibody-conjugated magnetic beads. The elution of the beads was then digested with trypsin. Then the samples were prepared using anion exchange solid-phase extraction before being detected by LC–MS/MS. The analytical performance of the method was validated in accordance with the recommendations of the Clinical and Laboratory Standards Institute C62-A and C64. Additionally, the method was compared to an immunofluorescent assay on the B.R.A.H.M.S Copeptin-proAVP KRYPTOR platform. A total of 221 residual serum samples were retrospectively collected from healthy individuals who underwent physical examinations and reference interval for copeptin was established. Diagnostic value of copeptin in identifying patients with diabetes insipidus was evaluated. Results A total of nine types of antibodies were obtained and their ability to capture serum copeptin was compared. The most effective antibody was subsequently utilized in this method after being conjugated to magnetic beads. The LC-MS/MS method had a total runtime of 8.5 minutes. Within-run precision ranged from 5.2 to 12.1%, between-run precision from 7.6 to 12.1%, and total variable coefficient from 8.1 to 13.5%. Copeptin quantitation showed linearity within the range of 5 to 1000 pg/mL, with a limit of detection at 2.5 pg/mL. Recovery rates ranged from 95.2% to 103.1%, and no significant matrix effect was observed with internal standard correction. The LC-MS/MS method had a good consistency with the immunoassay (r= 0.926, slope = 0.989). Additionally, the reference interval for copeptin was 3.66–58.25 pg/mL. A total of 44 patients diagnosed with central diabetes insipidus (CDI) were included and the diagnostic accuracy of copeptin for identifying CDI was evaluated by ROC curve analysis, yielding an area under the ROC curve of 0.922 (95% CI 0.882–0.952), with a sensitivity of 81.4% and a specificity of 89.7%. The cut-off value for CDI identification was determined to be 6.59 pg/mL. Conclusion We have demonstrated the accuracy and dependability of this LC-MS/MS method for quantifying copeptin. This innovative application showed satisfactory precision, a wide linear range, and a low limit of detection. Clinical studies are anticipated to be conducted to assess diagnostic accuracy using this method.

Authors

Institutions

Publication Details

Journal
Clinical Chemistry
Published
2025-10-01
DOI
https://doi.org/10.1093/clinchem/hvaf086.562
Primary Topic
Analytical Chemistry and Chromatography
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

B-169 A Novel Immunoprecipitation-based Liquid Chromatography–Tandem Mass Spectrometry Analysis for Accurate Determination for Copeptin in Human Serum

Danni Mu
Clinical Chemistry
Analytical Chemistry and Chromatography
article

B-169 A Novel Immunoprecipitation-based Liquid Chromatography–Tandem Mass Spectrometry Analysis for Accurate Determination for Copeptin in Human Serum

Danni Mu
article en

Abstract

Abstract Background Copeptin, a glycosylated peptide of 39 amino acids, is derived from the C-terminal segment of the arginine vasopressin (AVP) precursor peptide. It is released in an equimolar mode with AVP from the pituitary gland and has emerged as a stable and simple-to-measure surrogate marker for AVP. Copeptin demonstrates excellent clinical values, particularly in diagnosing polyuria-polydipsia syndromes such as diabetes insipidus. However, conventional immunoassay methods are limited and lack comparability. Therefore, our aim was to establish a liquid chromatography–tandem mass spectrometry (LC–MS/MS) method for quantifying copeptin levels in serum. Methods Copeptin was extracted from serum using immunoprecipitation with antibody-conjugated magnetic beads. The elution of the beads was then digested with trypsin. Then the samples were prepared using anion exchange solid-phase extraction before being detected by LC–MS/MS. The analytical performance of the method was validated in accordance with the recommendations of the Clinical and Laboratory Standards Institute C62-A and C64. Additionally, the method was compared to an immunofluorescent assay on the B.R.A.H.M.S Copeptin-proAVP KRYPTOR platform. A total of 221 residual serum samples were retrospectively collected from healthy individuals who underwent physical examinations and reference interval for copeptin was established. Diagnostic value of copeptin in identifying patients with diabetes insipidus was evaluated. Results A total of nine types of antibodies were obtained and their ability to capture serum copeptin was compared. The most effective antibody was subsequently utilized in this method after being conjugated to magnetic beads. The LC-MS/MS method had a total runtime of 8.5 minutes. Within-run precision ranged from 5.2 to 12.1%, between-run precision from 7.6 to 12.1%, and total variable coefficient from 8.1 to 13.5%. Copeptin quantitation showed linearity within the range of 5 to 1000 pg/mL, with a limit of detection at 2.5 pg/mL. Recovery rates ranged from 95.2% to 103.1%, and no significant matrix effect was observed with internal standard correction. The LC-MS/MS method had a good consistency with the immunoassay (r= 0.926, slope = 0.989). Additionally, the reference interval for copeptin was 3.66–58.25 pg/mL. A total of 44 patients diagnosed with central diabetes insipidus (CDI) were included and the diagnostic accuracy of copeptin for identifying CDI was evaluated by ROC curve analysis, yielding an area under the ROC curve of 0.922 (95% CI 0.882–0.952), with a sensitivity of 81.4% and a specificity of 89.7%. The cut-off value for CDI identification was determined to be 6.59 pg/mL. Conclusion We have demonstrated the accuracy and dependability of this LC-MS/MS method for quantifying copeptin. This innovative application showed satisfactory precision, a wide linear range, and a low limit of detection. Clinical studies are anticipated to be conducted to assess diagnostic accuracy using this method.

Clinical ChemistryVol. 71(Supplement_1)
Chinese Academy of Medical Sciences & Peking Union Medical College (CN), Peking Union Medical College Hospital (CN)
Openalex Percentile: Top 74%
Analytical Chemistry and Chromatography
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.