A-047 Evaluation of the Clinical Performance of the Thyroglobulin Assay on the Atellica IM Analyzer

Abstract Background Thyroglobulin (Tg) is a protein produced by thyroid follicular cells and functions as a prohormone in the intra-thyroid synthesis of thyroxine (T4) and triiodothyronine (T3). Lysosomes containing proteases cleave T4 and T3 from Tg, resulting in T4 and T3 release. Tg is present in the serum of healthy individuals and can be elevated in a number of disorders such as subacute thyroiditis, autonomous adenoma, amiodarone induced thyrotoxicosis and thyroid carcinoma. An important application of Tg measurement is the post-operative follow-up of patients with differentiated thyroid cancer (DTC). Quantitative Tg measurements are used as an aid in monitoring DTC patients who have undergone thyroidectomy, with or without radioiodine ablation. Here, studies were performed to evaluate reproducibility of the Atellica IM Tg assay on the Atellica IM analyzer and to evaluate assay clinical performance at a cutoff of 0.2 ng/mL. Values >0.2 ng/mL were deemed “positive” (evidence of possible cancer recurrence) as defined in 2015 ATA Guidelines. Methods Reproducibility testing was performed per CLSI EP05-A3, involving three levels of quality control (QC) material (0.221, 1.98, 25.8 ng/mL; 0.335, 3.00, 39.1 pmol/L) and a five-level patient sample pool (4.36, 44.0, 74.5, 120, 129 ng/mL; 6.61, 66.7, 113, 182, 195 pmol/L) tested in two runs per day for 5 days, three sites, three Atellica IM analyzers, and three Atellica IM Tg reagent lots. The same three reagent lots were tested at each of the three sites. Samples for clinical performance evaluation were prospectively collected from eligible subjects at sites in the United States. Testing was completed at one internal site. Primary analysis included 291 evaluable samples from 189 unique subjects. Assay result was compared to the status of structural disease as determined by cross-sectional or functional imaging results collected within 30 days of study blood draw. Clinical sensitivity, specificity, PPV and NPV were computed using a bootstrapping method to account for repeated Tg measurements within subjects. Results Quality control and patient sample pools recovered mean values between 0.221 and 129 ng/mL (0.335 and 195 pmol/L), spanning the assay range of 0.05-150 ng/mL (0.076-227.250 pmol/L) and covered the medical decision level of 0.2 ng/mL (0.3 pmol/L). Coefficients of Variation (CV) for reproducibility were =5.0% for samples recovering =1.000 ng/mL (1.5 pmol/L) and 9.0% for sample recovering <1.000 ng/mL. Sensitivity was 98.2% (95%CI 94.6%, 100.0%), specificity 53.4% (47.8%, 58.0%), PPV 10.0% (8.7%, 11.2%), and NPV 99.8% (99.5%, 100.0%) when assay result was compared to status of structural disease. Conclusion Overall, the Atellica IM Tg assay demonstrated acceptable clinical performance as an aid in the monitoring of DTC thyroidectomy patients with or without radioiodine ablation.

Authors

Institutions

Publication Details

Journal
Clinical Chemistry
Published
2025-10-01
DOI
https://doi.org/10.1093/clinchem/hvaf086.046
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A-047 Evaluation of the Clinical Performance of the Thyroglobulin Assay on the Atellica IM Analyzer

Nazira Özgen, Robert T. Singleton, Ryan Butterfield, Jennifer Snyder et al.
Clinical Chemistry
Clinical Laboratory Practices and Quality Control
article

A-047 Evaluation of the Clinical Performance of the Thyroglobulin Assay on the Atellica IM Analyzer

Nazira Özgen, Robert T. Singleton, Ryan Butterfield, Jennifer Snyder, Karoline Freeman, Lynn Schiavoni
article en

Abstract

Abstract Background Thyroglobulin (Tg) is a protein produced by thyroid follicular cells and functions as a prohormone in the intra-thyroid synthesis of thyroxine (T4) and triiodothyronine (T3). Lysosomes containing proteases cleave T4 and T3 from Tg, resulting in T4 and T3 release. Tg is present in the serum of healthy individuals and can be elevated in a number of disorders such as subacute thyroiditis, autonomous adenoma, amiodarone induced thyrotoxicosis and thyroid carcinoma. An important application of Tg measurement is the post-operative follow-up of patients with differentiated thyroid cancer (DTC). Quantitative Tg measurements are used as an aid in monitoring DTC patients who have undergone thyroidectomy, with or without radioiodine ablation. Here, studies were performed to evaluate reproducibility of the Atellica IM Tg assay on the Atellica IM analyzer and to evaluate assay clinical performance at a cutoff of 0.2 ng/mL. Values >0.2 ng/mL were deemed “positive” (evidence of possible cancer recurrence) as defined in 2015 ATA Guidelines. Methods Reproducibility testing was performed per CLSI EP05-A3, involving three levels of quality control (QC) material (0.221, 1.98, 25.8 ng/mL; 0.335, 3.00, 39.1 pmol/L) and a five-level patient sample pool (4.36, 44.0, 74.5, 120, 129 ng/mL; 6.61, 66.7, 113, 182, 195 pmol/L) tested in two runs per day for 5 days, three sites, three Atellica IM analyzers, and three Atellica IM Tg reagent lots. The same three reagent lots were tested at each of the three sites. Samples for clinical performance evaluation were prospectively collected from eligible subjects at sites in the United States. Testing was completed at one internal site. Primary analysis included 291 evaluable samples from 189 unique subjects. Assay result was compared to the status of structural disease as determined by cross-sectional or functional imaging results collected within 30 days of study blood draw. Clinical sensitivity, specificity, PPV and NPV were computed using a bootstrapping method to account for repeated Tg measurements within subjects. Results Quality control and patient sample pools recovered mean values between 0.221 and 129 ng/mL (0.335 and 195 pmol/L), spanning the assay range of 0.05-150 ng/mL (0.076-227.250 pmol/L) and covered the medical decision level of 0.2 ng/mL (0.3 pmol/L). Coefficients of Variation (CV) for reproducibility were =5.0% for samples recovering =1.000 ng/mL (1.5 pmol/L) and 9.0% for sample recovering <1.000 ng/mL. Sensitivity was 98.2% (95%CI 94.6%, 100.0%), specificity 53.4% (47.8%, 58.0%), PPV 10.0% (8.7%, 11.2%), and NPV 99.8% (99.5%, 100.0%) when assay result was compared to status of structural disease. Conclusion Overall, the Atellica IM Tg assay demonstrated acceptable clinical performance as an aid in the monitoring of DTC thyroidectomy patients with or without radioiodine ablation.

Clinical ChemistryVol. 71(Supplement_1)
Siemens Healthcare (United States) (US), Siemens Healthineers (Germany) (DE), Mental Health Association of Westchester County (US)
Openalex Percentile: Top 66%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.