A-125 Evaluation of Commercial Serum Indices Quality Control Materials for a Coagulation HIL Module

Abstract Background The STA R MAX^3 and STA Compact Max^3, coagulation analyzers developed by Diagnostica Stago (Parsippany, NJ), are able to perform clotting, chromogenic and immunoturbidimetric tests simultaneously, using a Viscosity Based Detection System. They are also equipped with a preanalytical module that can evaluate for the presence of hemolysis, icterus, and lipemia (HIL) in samples prior to analysis. Evaluation of specimen suitability for downstream analytical testing and the identification of potential interferents in coagulation specimens is essential for ensuring the accuracy and reliability of clinical results. In the clinical laboratory, hemolysis, icterus, and lipemia are evaluated for coagulation testing using spectrophotometric methods. While coagulation laboratories rely on analyte-specific quality control (QC) materials to monitor test and instrument performance, quality assurance materials evaluating specimen integrity checks are infrequently utilized. The purpose of this evaluation was to assess a commercial serum indices product designed to monitor an instrument*s performance in detecting hemolyzed, icteric, or lipemic samples. Methods Commercially available specimen integrity materials were evaluated on two coagulation analyzers: the STA R MAX^3 and STA Compact Max^3 by a certified and trained medical laboratory professional. Target arbitrary values for the hemolysis, icterus, and lipemia QC materials were 200, 23, and 184, respectively. Each serum indices level was tested in duplicate, twice a day, across both analyzers for with the same aliquots for each analyzer and for each interferent. Semi-quantitative data comparisons were performed using a proposed lookup table for the preanalytical module that compared Stago HIL indices to Bio-Rad Liquichek (Bio-Rad Laboratories, Irvine, CA) reference tables. Results Between the STA R MAX^3 and STA Compact Max^3, hemolysis and icterus were determined to have the same semi-quantitative acceptability thresholds as those listed in the manufacturer package insert using the proposed lookup table. Lipemia testing, based on human sourced material, resulted in a new acceptability threshold that deviated slightly from manufacturer recommendations (1 semi-quantitative level higher for the lipemia material). Manufacturer provided quantitative values obtained for the Bio-Rad Liquichek Serum Indices product were approximate targets from a clinical chemistry analyzer. Conclusion The Bio-Rad Liquichek Serum Indices product can serve as an effective means of monitoring specimen integrity checks on the STA R MAX3 and STA Compact Max3 EPC module and brings benefit to the total quality assurance process. Independent evaluation of commercial HIL indices material as an unassayed product is recommended to enable informative implementation of specimen quality criteria and to provide guidance to laboratorians and providers on the nature of these interferences.

Authors

Institutions

Publication Details

Journal
Clinical Chemistry
Published
2025-10-01
DOI
https://doi.org/10.1093/clinchem/hvaf086.121
Primary Topic
Blood donation and transfusion practices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A-125 Evaluation of Commercial Serum Indices Quality Control Materials for a Coagulation HIL Module

LaShanta Brice, Jennifer Blockhaus
Clinical Chemistry
Blood donation and transfusion practices
article

A-125 Evaluation of Commercial Serum Indices Quality Control Materials for a Coagulation HIL Module

LaShanta Brice, Jennifer Blockhaus
article en

Abstract

Abstract Background The STA R MAX^3 and STA Compact Max^3, coagulation analyzers developed by Diagnostica Stago (Parsippany, NJ), are able to perform clotting, chromogenic and immunoturbidimetric tests simultaneously, using a Viscosity Based Detection System. They are also equipped with a preanalytical module that can evaluate for the presence of hemolysis, icterus, and lipemia (HIL) in samples prior to analysis. Evaluation of specimen suitability for downstream analytical testing and the identification of potential interferents in coagulation specimens is essential for ensuring the accuracy and reliability of clinical results. In the clinical laboratory, hemolysis, icterus, and lipemia are evaluated for coagulation testing using spectrophotometric methods. While coagulation laboratories rely on analyte-specific quality control (QC) materials to monitor test and instrument performance, quality assurance materials evaluating specimen integrity checks are infrequently utilized. The purpose of this evaluation was to assess a commercial serum indices product designed to monitor an instrument*s performance in detecting hemolyzed, icteric, or lipemic samples. Methods Commercially available specimen integrity materials were evaluated on two coagulation analyzers: the STA R MAX^3 and STA Compact Max^3 by a certified and trained medical laboratory professional. Target arbitrary values for the hemolysis, icterus, and lipemia QC materials were 200, 23, and 184, respectively. Each serum indices level was tested in duplicate, twice a day, across both analyzers for with the same aliquots for each analyzer and for each interferent. Semi-quantitative data comparisons were performed using a proposed lookup table for the preanalytical module that compared Stago HIL indices to Bio-Rad Liquichek (Bio-Rad Laboratories, Irvine, CA) reference tables. Results Between the STA R MAX^3 and STA Compact Max^3, hemolysis and icterus were determined to have the same semi-quantitative acceptability thresholds as those listed in the manufacturer package insert using the proposed lookup table. Lipemia testing, based on human sourced material, resulted in a new acceptability threshold that deviated slightly from manufacturer recommendations (1 semi-quantitative level higher for the lipemia material). Manufacturer provided quantitative values obtained for the Bio-Rad Liquichek Serum Indices product were approximate targets from a clinical chemistry analyzer. Conclusion The Bio-Rad Liquichek Serum Indices product can serve as an effective means of monitoring specimen integrity checks on the STA R MAX3 and STA Compact Max3 EPC module and brings benefit to the total quality assurance process. Independent evaluation of commercial HIL indices material as an unassayed product is recommended to enable informative implementation of specimen quality criteria and to provide guidance to laboratorians and providers on the nature of these interferences.

Clinical ChemistryVol. 71(Supplement_1)
Reckitt Benckiser (United States) (US)
Openalex Percentile: Top 64%
Blood donation and transfusion practices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.