Abstract Background Liver enzyme assays, including alanine aminotransferase (ALT), aspartate aminotransferase (AST), gamma-glutamyltransferase (GGT), and alkaline phosphatase (ALP), are essential in clinical laboratories for assessing liver function, diagnosing liver diseases, and monitoring treatment efficacy. Thus, accurate measurement of these enzyme activities is crucial for clinical diagnostics. Recently, Abbott introduced a next generation of Alinity and Architect enzyme assays, incorporating a multianalyte calibrator (Consolidated Chemistry Calibrator, ConCC) with extended stability. For these new formulation assays, both ConCC and calibration factors are offered for all the assays except GGT, which had better performance with the ConCC calibrator. These assays achieve high Sigma metrics on the Alinity c system, minimizing variability and errors. This study evaluated the analytical performance of these newly developed next generation assays on the Abbott Alinity c platform. Methods Using 2 levels of quality control (QC) material (Bio-Rad Chemistry UA) and 3 pooled patient samples, we assessed imprecision by measuring these 5 samples twice per day (morning and afternoon), for five days. Acceptable imprecision and bias were determined based on the Accreditation Canada Diagnostics (ACDx) recommendations. Linearity testing consisted of 6 levels of commercially available linearity materials, with 3 replicates per level. Method comparison between the next generation assays and on-market conventional assays were evaluated in duplicates using patient specimens (n= 145 – 155) on the Abbott Alinity c platform. Passing-Bablok and Bland-Altman plots were used for the method comparison analyses. The precision and bias (external standard) studies were used to calculate the Sigma-metric using the formula, Sigma-metric=(%TEa-|%bias|)/%CV. TEa was defined based on ACDx and Clinical Laboratory Improvement Amendments (CLIA) recommendations. Results The Alinity next generation assays demonstrated acceptable imprecision, meeting the ACDx goals of ±2 U/L for concentrations =40 U/L and ±4% for concentrations >40 U/L for AST, ALT and GGT, and for ALP at ±4 U/L for concentrations =100 U/L and ±4% for concentrations >100 U/L. Furthermore, these assays exhibited linearity across the six concentration levels tested. All next generation Alinity clinical chemistry enzyme assays showed a Pearson*s R value of 1.0, indicating a strong linear correlation. The linearity slope ranged from 0.76 (ALP2) to 1.04 (AST2) whereas the y-intercept ranged from –59.40 (GGT2) to 8.78 (AST2). The majority of next generation assays performed at or above 6 Sigma. Good agreements were observed between the on-market assay and the next generation (ConCC and factor calibrated) assays. Conclusion The Alinity next generation assays (ALT2, AST2, GGT2 and ALP2) demonstrated acceptable performance for precision and linearity, with good agreement with the conventional factor-based assays on the Alinity c system. These next generation assays had a high sigma value; hence laboratories can expect excellent performance. The precision and method comparison agreement with the conventional assays were satisfactory.

Authors

Institutions

Publication Details

Journal
Clinical Chemistry
Published
2025-10-01
DOI
https://doi.org/10.1093/clinchem/hvaf086.117
Primary Topic
Fungal Plant Pathogen Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A-121 Analytical performance evaluation of the next generation Abbott enzyme assays on the Alinity c system

Pow Lee Cheng, Vathany Kulasingam, Marvin Berman, Meshach Asare-Werehene et al.
Clinical Chemistry
Fungal Plant Pathogen Control
article

A-121 Analytical performance evaluation of the next generation Abbott enzyme assays on the Alinity c system

Pow Lee Cheng, Vathany Kulasingam, Marvin Berman, Meshach Asare-Werehene, Xiaoyan Wang
article en

Abstract

Abstract Background Liver enzyme assays, including alanine aminotransferase (ALT), aspartate aminotransferase (AST), gamma-glutamyltransferase (GGT), and alkaline phosphatase (ALP), are essential in clinical laboratories for assessing liver function, diagnosing liver diseases, and monitoring treatment efficacy. Thus, accurate measurement of these enzyme activities is crucial for clinical diagnostics. Recently, Abbott introduced a next generation of Alinity and Architect enzyme assays, incorporating a multianalyte calibrator (Consolidated Chemistry Calibrator, ConCC) with extended stability. For these new formulation assays, both ConCC and calibration factors are offered for all the assays except GGT, which had better performance with the ConCC calibrator. These assays achieve high Sigma metrics on the Alinity c system, minimizing variability and errors. This study evaluated the analytical performance of these newly developed next generation assays on the Abbott Alinity c platform. Methods Using 2 levels of quality control (QC) material (Bio-Rad Chemistry UA) and 3 pooled patient samples, we assessed imprecision by measuring these 5 samples twice per day (morning and afternoon), for five days. Acceptable imprecision and bias were determined based on the Accreditation Canada Diagnostics (ACDx) recommendations. Linearity testing consisted of 6 levels of commercially available linearity materials, with 3 replicates per level. Method comparison between the next generation assays and on-market conventional assays were evaluated in duplicates using patient specimens (n= 145 – 155) on the Abbott Alinity c platform. Passing-Bablok and Bland-Altman plots were used for the method comparison analyses. The precision and bias (external standard) studies were used to calculate the Sigma-metric using the formula, Sigma-metric=(%TEa-|%bias|)/%CV. TEa was defined based on ACDx and Clinical Laboratory Improvement Amendments (CLIA) recommendations. Results The Alinity next generation assays demonstrated acceptable imprecision, meeting the ACDx goals of ±2 U/L for concentrations =40 U/L and ±4% for concentrations >40 U/L for AST, ALT and GGT, and for ALP at ±4 U/L for concentrations =100 U/L and ±4% for concentrations >100 U/L. Furthermore, these assays exhibited linearity across the six concentration levels tested. All next generation Alinity clinical chemistry enzyme assays showed a Pearson*s R value of 1.0, indicating a strong linear correlation. The linearity slope ranged from 0.76 (ALP2) to 1.04 (AST2) whereas the y-intercept ranged from –59.40 (GGT2) to 8.78 (AST2). The majority of next generation assays performed at or above 6 Sigma. Good agreements were observed between the on-market assay and the next generation (ConCC and factor calibrated) assays. Conclusion The Alinity next generation assays (ALT2, AST2, GGT2 and ALP2) demonstrated acceptable performance for precision and linearity, with good agreement with the conventional factor-based assays on the Alinity c system. These next generation assays had a high sigma value; hence laboratories can expect excellent performance. The precision and method comparison agreement with the conventional assays were satisfactory.

Clinical ChemistryVol. 71(Supplement_1)
University Health Network (CA), University of Toronto (CA), Toronto General Hospital (CA), Abbott (United States) (US)
Openalex Percentile: Top 78%
Fungal Plant Pathogen Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.