A-201 Development of a Predictive Model for Hemolysis Interference on Blood Gas Analysis

Abstract Background Hemolysis is one of the important preanalytical factor that can influence the laboratory test results. As arterial blood gas analysis (ABGA) is performed with whole blood, it is difficult to visually check hemolysis state of the specimen, and even blood gas analyzers cannot detect hemolysis. However, there is not enough consensus or guideline on which parameters are influenced with hemolysis. This study was aimed to identify the influence of hemolysis on ABGA results and develop a predictive model for hemolysis interference. Methods A total of 142 residual arterial blood specimens were collected from a teritary hospital in South Korea. Samples were aliquoted into three groups for mechanical hemolysis. Hemolysis was induced via 16, 22, and 26 Gauge needles and measured using the Profile pHOx Ultra Blood Gas Analyzer (Nova Biomedical, Waltham, MA, USA). The remaining blood was centrifuged, and hemolysis index was identified using plasma. The differences among groups were evaluated by one-way analysis of variance (ANOVA). In addition, the test results were divided into two sets: the training set and the test set. R studio (version 4.3.1) was used for the statistical analysis and the development of the predictive model. To select the optimal combination of parameters, we employed the backward elimination method. Results Among the parameters, pH and K values increased whereas pCO2, Na, Ca2+, and HCO3- values decreased. Values of pCO2, Hb, K, and Ca2+ increased as the degree of hemolysis increased with % biases exceeding the desirable bias. When the predictive model developed with training set was tested, the model that included pCO2, Hb, K, Cl, Ca2+, Lac, and tCO2 as variables demonstrated the top-flight performance. This model achieved the highest area under the curve (AUC) among all tested combinations (AUC=0.905). It exhibited a sensitivity of 89.6% and a specificity of 92.1%. Conclusion This study confirms that hemolysis influences the values of pH, pCO2, and K significantly. Furthermore, a predictive model was validated for detection of hemolysis interference in whole blood specimens used for ABGA. Therefore, the predictive model to detect spurious hemolysis should be adopted in ABGA instruments in clinical laboratories and POCT environments to reduce the analytical errors caused by hemolysis.

Authors

Institutions

Publication Details

Journal
Clinical Chemistry
Published
2025-10-01
DOI
https://doi.org/10.1093/clinchem/hvaf086.195
Primary Topic
Metabolomics and Mass Spectrometry Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A-201 Development of a Predictive Model for Hemolysis Interference on Blood Gas Analysis

Hanmil Jang, Jong‐Baeck Lim, Sang‐Guk Lee, Hyein Kang et al.
Clinical Chemistry
Metabolomics and Mass Spectrometry Studies
article

A-201 Development of a Predictive Model for Hemolysis Interference on Blood Gas Analysis

Hanmil Jang, Jong‐Baeck Lim, Sang‐Guk Lee, Hyein Kang, John Hoon Rim
article en

Abstract

Abstract Background Hemolysis is one of the important preanalytical factor that can influence the laboratory test results. As arterial blood gas analysis (ABGA) is performed with whole blood, it is difficult to visually check hemolysis state of the specimen, and even blood gas analyzers cannot detect hemolysis. However, there is not enough consensus or guideline on which parameters are influenced with hemolysis. This study was aimed to identify the influence of hemolysis on ABGA results and develop a predictive model for hemolysis interference. Methods A total of 142 residual arterial blood specimens were collected from a teritary hospital in South Korea. Samples were aliquoted into three groups for mechanical hemolysis. Hemolysis was induced via 16, 22, and 26 Gauge needles and measured using the Profile pHOx Ultra Blood Gas Analyzer (Nova Biomedical, Waltham, MA, USA). The remaining blood was centrifuged, and hemolysis index was identified using plasma. The differences among groups were evaluated by one-way analysis of variance (ANOVA). In addition, the test results were divided into two sets: the training set and the test set. R studio (version 4.3.1) was used for the statistical analysis and the development of the predictive model. To select the optimal combination of parameters, we employed the backward elimination method. Results Among the parameters, pH and K values increased whereas pCO2, Na, Ca2+, and HCO3- values decreased. Values of pCO2, Hb, K, and Ca2+ increased as the degree of hemolysis increased with % biases exceeding the desirable bias. When the predictive model developed with training set was tested, the model that included pCO2, Hb, K, Cl, Ca2+, Lac, and tCO2 as variables demonstrated the top-flight performance. This model achieved the highest area under the curve (AUC) among all tested combinations (AUC=0.905). It exhibited a sensitivity of 89.6% and a specificity of 92.1%. Conclusion This study confirms that hemolysis influences the values of pH, pCO2, and K significantly. Furthermore, a predictive model was validated for detection of hemolysis interference in whole blood specimens used for ABGA. Therefore, the predictive model to detect spurious hemolysis should be adopted in ABGA instruments in clinical laboratories and POCT environments to reduce the analytical errors caused by hemolysis.

Clinical ChemistryVol. 71(Supplement_1)
Yonsei University (KR), Keimyung University (KR)
Openalex Percentile: Top 76%
Metabolomics and Mass Spectrometry Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.