B-130 Quantification of the Sensitive Inflammatory Marker Serum Amyloid A (SAA) by Electrochemiluminescence Immunoassay

Abstract Background Serum Amyloid A (SAA) is an acute phase response (APR) protein. As a marker of both acute and chronic inflammation. SAA has been used to investigate common systemic autoimmune diseases and infections like SARS-CoV-2 as well as less common inflammatory conditions like familial Mediterranean fever (FMF) and Behcet’s disease. Similar to but more sensitive than C-reactive protein (CRP), SAA levels rise quickly more than 100-fold during episodes of inflammation and then fall as inflammation resolves. SAA is also the soluble precursor of Amyloid A which can accumulate in tissues such as heart, kidney, liver, spleen, and nervous system as amyloidosis. Amyloidosis affects FMF and Behcet’s patients who have not only episodic inflammation but also prolonged subclinical inflammation. Importantly, monitoring SAA levels allows for the detection of subclinical inflammation, and better control of subclinical inflammation holds the promise of long-term benefits in growth, intellectual function, cardiovascular health and reduced amyloidosis with reduced end-stage kidney disease. Here, we developed an electrochemiluminescence immunoassay (ECLIA) to quantify SAA with robust analytical performance for such diverse clinical applications as evaluation and monitoring of patients with chronic inflammatory diseases. Methods Detection and quantification of SAA was accomplished with a lab developed test (LDT) two-step ECLIA. The analytical performance of the assay was validated for the parameters of accuracy, precision, lower limit of quantitation (LLOQ), upper limit of quantitation (ULOQ), and dilutional linearity. We established a reference interval cut point using serum samples from >120 individuals with no history of inflammatory disease and no recent evidence of inflammation, infection, or injury. Calculation of reference interval cut point was determined using a transformed parametric test. Results The analytical measurement range (AMR) of the assay was determined to be from 0.4 µg/mL to 625 µg/mL. Samples diluted linearly >1,000-fold (recovery within 80 – 120%) allowing for a maximum reportable value of 625,000 µg/mL and a lower limit of quantitation (LLOQ) of 0.4 µg/mL. Within the AMR, standard bias was = ±8.5% and standard imprecision was = 9.1% CV. Sample imprecision was = 10.9% CV. The reference interval cut point was determined to be < 5 µg/mL (95% CI, 3.78 – 5.51) with a confidence interval of 0.21. Of the 120 samples screened, < 15% fell below the assay LLOQ and < 1.7% were above the assigned reference interval cut point. Conclusion SAA is a highly sensitive marker of both acute and chronic inflammation. Levels of SAA in serum have clinical utility in identifying inflammation, especially subclinical inflammation, in FMF and other diseases that lack other more specific biomarkers. We report our validation of this highly sensitive LDT electrochemiluminescence assay to quantify SAA with the necessary analytic range and precision to allow proactive management of diverse inflammatory conditions aimed toward amelioration of their long-term sequelae.

Authors

Institutions

Publication Details

Journal
Clinical Chemistry
Published
2025-10-01
DOI
https://doi.org/10.1093/clinchem/hvaf086.526
Primary Topic
Extracellular vesicles in disease
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

B-130 Quantification of the Sensitive Inflammatory Marker Serum Amyloid A (SAA) by Electrochemiluminescence Immunoassay

Kelly Chun, Monique Bastidas, Jane Yang, Brandon Bauer
Clinical Chemistry
Extracellular vesicles in disease
article

B-130 Quantification of the Sensitive Inflammatory Marker Serum Amyloid A (SAA) by Electrochemiluminescence Immunoassay

Kelly Chun, Monique Bastidas, Jane Yang, Brandon Bauer
article en

Abstract

Abstract Background Serum Amyloid A (SAA) is an acute phase response (APR) protein. As a marker of both acute and chronic inflammation. SAA has been used to investigate common systemic autoimmune diseases and infections like SARS-CoV-2 as well as less common inflammatory conditions like familial Mediterranean fever (FMF) and Behcet’s disease. Similar to but more sensitive than C-reactive protein (CRP), SAA levels rise quickly more than 100-fold during episodes of inflammation and then fall as inflammation resolves. SAA is also the soluble precursor of Amyloid A which can accumulate in tissues such as heart, kidney, liver, spleen, and nervous system as amyloidosis. Amyloidosis affects FMF and Behcet’s patients who have not only episodic inflammation but also prolonged subclinical inflammation. Importantly, monitoring SAA levels allows for the detection of subclinical inflammation, and better control of subclinical inflammation holds the promise of long-term benefits in growth, intellectual function, cardiovascular health and reduced amyloidosis with reduced end-stage kidney disease. Here, we developed an electrochemiluminescence immunoassay (ECLIA) to quantify SAA with robust analytical performance for such diverse clinical applications as evaluation and monitoring of patients with chronic inflammatory diseases. Methods Detection and quantification of SAA was accomplished with a lab developed test (LDT) two-step ECLIA. The analytical performance of the assay was validated for the parameters of accuracy, precision, lower limit of quantitation (LLOQ), upper limit of quantitation (ULOQ), and dilutional linearity. We established a reference interval cut point using serum samples from >120 individuals with no history of inflammatory disease and no recent evidence of inflammation, infection, or injury. Calculation of reference interval cut point was determined using a transformed parametric test. Results The analytical measurement range (AMR) of the assay was determined to be from 0.4 µg/mL to 625 µg/mL. Samples diluted linearly >1,000-fold (recovery within 80 – 120%) allowing for a maximum reportable value of 625,000 µg/mL and a lower limit of quantitation (LLOQ) of 0.4 µg/mL. Within the AMR, standard bias was = ±8.5% and standard imprecision was = 9.1% CV. Sample imprecision was = 10.9% CV. The reference interval cut point was determined to be < 5 µg/mL (95% CI, 3.78 – 5.51) with a confidence interval of 0.21. Of the 120 samples screened, < 15% fell below the assay LLOQ and < 1.7% were above the assigned reference interval cut point. Conclusion SAA is a highly sensitive marker of both acute and chronic inflammation. Levels of SAA in serum have clinical utility in identifying inflammation, especially subclinical inflammation, in FMF and other diseases that lack other more specific biomarkers. We report our validation of this highly sensitive LDT electrochemiluminescence assay to quantify SAA with the necessary analytic range and precision to allow proactive management of diverse inflammatory conditions aimed toward amelioration of their long-term sequelae.

Clinical ChemistryVol. 71(Supplement_1)
LabCorp (United States) (US), Optical Physics Company (United States) (US), StereoVision Imaging (United States) (US)
Openalex Percentile: Top 76%
Extracellular vesicles in disease
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.