B-282 Validation of a Liquid Chromatography-Tandem Mass Spectrometry Method for Simultaneous Quantitation of Voriconazole, Posaconazole, Itraconazole, and Hydroxyitraconazole in Human Plasma

Abstract Background Azole-based antifungals are the first-line therapy for many common mycoses and are increasingly used prophylactically to protect immunocompromised patients. Due to variability in their metabolism and bioavailability, therapeutic drug monitoring is crucial to avoid toxicity while ensuring maximum efficacy. Methods Sample preparation involved protein precipitation with a methanol and internal standard mixture, followed by centrifugation. Isotope-labeled analogues for each analyte served as internal standards. Chromatographic separation was performed using an ACQUITY UPLC BEH C18 Column (30 mm x 2.1 mm, 1.7 µm) with a BEH C18 VanGuard Pre-column (2.1 mm x 5 mm, 1.7 µm) on an ACQUITY UPLC I-Class System. Gradient elution was achieved with Mobile phase A and B consisting of 2 mM ammonium acetate in 0.1% formic acid in H2O and methanol, respectively, at a flow rate of 0.8 mL/min. Azoles were detected using a triple-stage quadrupole mass spectrometer in selected reaction monitoring mode with positive heated electrospray ionization, within a runtime of 3.00 minutes. Results Linearity and calibration ranges for all azoles were verified from 0.05 to 10 µg/mL, with a limit of quantitation at 0.05 µg/mL. Within-run and between-run precisions ranged from 0.6% to 6.7% for all azoles. Accuracy was assessed by comparing results with a reference lab, showing high correlation with an average slope of 1.06, intercept of 0.06, and R value of 0.99. Comparison between two instrument systems also demonstrated similar results, with an R value of 0.99. Dilution was verified by reduced volume injection at a 1:2 dilution, extending reportable ranges up to 20 µg/mL. Carryover studies, conducted by interspersing high concentration and negative samples, passed the EP Evaluator protocol. Matrix-induced ion suppression/enhancement studies revealed no interference from the EDTA plasma matrix, except for hydroxyitraconazole. Consequently, hydroxyitraconazole results are not reported when itraconazole is undetectable, as hydroxyitraconazole was enhanced by over 30% of the spiked concentration due to the sample matrix. Whole blood stability was assessed at 8 hours at room temperature before centrifugation, with results acceptable based on a 20% total allowable error. Conclusion A simple yet reliable liquid chromatography-tandem mass spectrometry method was validated that enables quantification of azoles in EDTA plasma across therapeutic ranges. These new assays have significantly improved result turnaround time and have been successfully implemented in our therapeutic drug monitoring service, making them suitable for routine monitoring of antifungal therapy, especially in severely ill patients.

Authors

Institutions

Publication Details

Journal
Clinical Chemistry
Published
2025-10-01
DOI
https://doi.org/10.1093/clinchem/hvaf086.669
Primary Topic
Parasitic Infections and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

B-282 Validation of a Liquid Chromatography-Tandem Mass Spectrometry Method for Simultaneous Quantitation of Voriconazole, Posaconazole, Itraconazole, and Hydroxyitraconazole in Human Plasma

R. Rodney Howell, Shu‐Ling Fan
Clinical Chemistry
Parasitic Infections and Diagnostics
article

B-282 Validation of a Liquid Chromatography-Tandem Mass Spectrometry Method for Simultaneous Quantitation of Voriconazole, Posaconazole, Itraconazole, and Hydroxyitraconazole in Human Plasma

R. Rodney Howell, Shu‐Ling Fan
article en

Abstract

Abstract Background Azole-based antifungals are the first-line therapy for many common mycoses and are increasingly used prophylactically to protect immunocompromised patients. Due to variability in their metabolism and bioavailability, therapeutic drug monitoring is crucial to avoid toxicity while ensuring maximum efficacy. Methods Sample preparation involved protein precipitation with a methanol and internal standard mixture, followed by centrifugation. Isotope-labeled analogues for each analyte served as internal standards. Chromatographic separation was performed using an ACQUITY UPLC BEH C18 Column (30 mm x 2.1 mm, 1.7 µm) with a BEH C18 VanGuard Pre-column (2.1 mm x 5 mm, 1.7 µm) on an ACQUITY UPLC I-Class System. Gradient elution was achieved with Mobile phase A and B consisting of 2 mM ammonium acetate in 0.1% formic acid in H2O and methanol, respectively, at a flow rate of 0.8 mL/min. Azoles were detected using a triple-stage quadrupole mass spectrometer in selected reaction monitoring mode with positive heated electrospray ionization, within a runtime of 3.00 minutes. Results Linearity and calibration ranges for all azoles were verified from 0.05 to 10 µg/mL, with a limit of quantitation at 0.05 µg/mL. Within-run and between-run precisions ranged from 0.6% to 6.7% for all azoles. Accuracy was assessed by comparing results with a reference lab, showing high correlation with an average slope of 1.06, intercept of 0.06, and R value of 0.99. Comparison between two instrument systems also demonstrated similar results, with an R value of 0.99. Dilution was verified by reduced volume injection at a 1:2 dilution, extending reportable ranges up to 20 µg/mL. Carryover studies, conducted by interspersing high concentration and negative samples, passed the EP Evaluator protocol. Matrix-induced ion suppression/enhancement studies revealed no interference from the EDTA plasma matrix, except for hydroxyitraconazole. Consequently, hydroxyitraconazole results are not reported when itraconazole is undetectable, as hydroxyitraconazole was enhanced by over 30% of the spiked concentration due to the sample matrix. Whole blood stability was assessed at 8 hours at room temperature before centrifugation, with results acceptable based on a 20% total allowable error. Conclusion A simple yet reliable liquid chromatography-tandem mass spectrometry method was validated that enables quantification of azoles in EDTA plasma across therapeutic ranges. These new assays have significantly improved result turnaround time and have been successfully implemented in our therapeutic drug monitoring service, making them suitable for routine monitoring of antifungal therapy, especially in severely ill patients.

Clinical ChemistryVol. 71(Supplement_1)
Tennessee Department of Health (US), Tennessee Oncology (US), Nashville Oncology Associates (US), Vanderbilt University Medical Center (US)
Openalex Percentile: Top 72%
Parasitic Infections and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.