Long-term performance of a low-modulus acrylic cement in a simulated physiological environment

A vertebroplasty bone cement, designed to exhibit mechanical properties closer to those of the vertebral bone, was investigated. Indeed, adjacent fractures are attributed to an excessive difference in elastic modulus between the vertebrae and the bone cement implying another operation to repair the newly damaged vertebra. The proposed solution to lower the mechanical properties is the introduction of gelatine in the powder of the cement. This low modulus bone cement was studied by quasi-static compression, dynamic mechanical analysis and fatigue tests. Compression testing showed a significant decrease of the elastic modulus of 33 % after 6 months of immersion in a PBS solution at 37 °C, confirmed by the dynamic mechanical analysis. This analysis also showed that immersion and the introduction of gelatine in the cement influence the relaxations of the PMMA matrix. Fatigue performance was also studied with and without gelatine and showed that it presents a characteristic life of at least 24,600 cycles for a load of 7 MPa. While compression testing indicates a reduction in mechanical performance, all others tests demonstrated that gelatine does not significantly impair the overall properties of acrylic bone cement that still possess sufficient properties to achieve its main goal to stabilise damaged vertebrae.

Authors

Institutions

Publication Details

Journal
Medical Engineering & Physics
Published
2025-09-29
DOI
https://doi.org/10.1016/j.medengphy.2025.104447
Primary Topic
Spinal Fractures and Fixation Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Long-term performance of a low-modulus acrylic cement in a simulated physiological environment

C. Lavergne, Marcq Fabien, C. Lacabanne, Eric Dantras A et al.
Medical Engineering & Physics
Spinal Fractures and Fixation Techniques
article

Long-term performance of a low-modulus acrylic cement in a simulated physiological environment

C. Lavergne, Marcq Fabien, C. Lacabanne, Eric Dantras A, C Lardilleux, M Faruch
article en

Abstract

A vertebroplasty bone cement, designed to exhibit mechanical properties closer to those of the vertebral bone, was investigated. Indeed, adjacent fractures are attributed to an excessive difference in elastic modulus between the vertebrae and the bone cement implying another operation to repair the newly damaged vertebra. The proposed solution to lower the mechanical properties is the introduction of gelatine in the powder of the cement. This low modulus bone cement was studied by quasi-static compression, dynamic mechanical analysis and fatigue tests. Compression testing showed a significant decrease of the elastic modulus of 33 % after 6 months of immersion in a PBS solution at 37 °C, confirmed by the dynamic mechanical analysis. This analysis also showed that immersion and the introduction of gelatine in the cement influence the relaxations of the PMMA matrix. Fatigue performance was also studied with and without gelatine and showed that it presents a characteristic life of at least 24,600 cycles for a load of 7 MPa. While compression testing indicates a reduction in mechanical performance, all others tests demonstrated that gelatine does not significantly impair the overall properties of acrylic bone cement that still possess sufficient properties to achieve its main goal to stabilise damaged vertebrae.

Medical Engineering & PhysicsVol. 145(1)
Centre National de la Recherche Scientifique (FR), Université Fédérale de Toulouse Midi-Pyrénées (FR), Centre Hospitalier Universitaire de Toulouse (FR), Centre Interuniversitaire de Recherche et d’Ingénierie des Matériaux (FR), Teknimed (France) (FR)
Openalex Percentile: Top 57%
Spinal Fractures and Fixation Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.