Approach to Optimization of Mass Spectrometric Detection Parameters for Identificationof Ultra-Small Amounts of Highly Toxic Substances

The procedure of optimization of the following parameters is described: scanning speed, duration of registration of selective transitions (dwell time), duration of delay in transition from one selective transition to another (pause time). A chromatograph for HPLC with a three-quadrupole mass spectrometric detector (LC-MS/MS Shimadzu 8050) was used. Chromatographic separation was carried out on a column with reversed-phase sorbent Phenomenex Kinetex C18. As mobile phase A was used 0.1 % solution of formic acid in water with addition of 10 mM/L ammonium formate, as mobile phase B – 0.1 % solution of formic acid in methanol with addition of 10 mmol/L ammonium formate. The optimal parameters for confirmatory methods were established: duration of registration of selective transitions – not less than 10 ms, duration of delay in transition from one selective transition to another – 1 ms, scanning speed – from 1000 to 5000 scans per second. This technique has been successfully applied in routine chemical and toxicological studies of urine samples with low content of various narcotic and medicinal substances.

Authors

Publication Details

Journal
Журнал аналитической химии / Journal of Analytical Chemistry
Published
2025-01-01
DOI
https://doi.org/10.7868/s3034512x25020043
Primary Topic
Advanced Chemical Sensor Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Approach to Optimization of Mass Spectrometric Detection Parameters for Identificationof Ultra-Small Amounts of Highly Toxic Substances

M. Sh. Aigumov
Журнал аналитической химии / Journal of Analytical Chemistry
Advanced Chemical Sensor Technologies
article

Approach to Optimization of Mass Spectrometric Detection Parameters for Identificationof Ultra-Small Amounts of Highly Toxic Substances

M. Sh. Aigumov
article en

Abstract

The procedure of optimization of the following parameters is described: scanning speed, duration of registration of selective transitions (dwell time), duration of delay in transition from one selective transition to another (pause time). A chromatograph for HPLC with a three-quadrupole mass spectrometric detector (LC-MS/MS Shimadzu 8050) was used. Chromatographic separation was carried out on a column with reversed-phase sorbent Phenomenex Kinetex C18. As mobile phase A was used 0.1 % solution of formic acid in water with addition of 10 mM/L ammonium formate, as mobile phase B – 0.1 % solution of formic acid in methanol with addition of 10 mmol/L ammonium formate. The optimal parameters for confirmatory methods were established: duration of registration of selective transitions – not less than 10 ms, duration of delay in transition from one selective transition to another – 1 ms, scanning speed – from 1000 to 5000 scans per second. This technique has been successfully applied in routine chemical and toxicological studies of urine samples with low content of various narcotic and medicinal substances.

Журнал аналитической химии / Journal of Analytical ChemistryVol. 80(2)
Openalex Percentile: Top 74%
Advanced Chemical Sensor Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.