Design and verification of the testing device for thoracic aortic stent grafts

Objective: To design a testing device for measuring the radial support force and bending spring back force of stent grafts and evaluate its effectiveness. Methods: A radial force and spring-back force testing device was designed to integrate with a tensile testing machine. The radial compression and bending characteristics of stent grafts for thoracic aorta applications were analyzed, and the corresponding conversion formulas were derived. A custom stent ring fixture was fabricated, and a five-wave gradient stent was sewn. Both physical experiments and finite element simulations were conducted. Radial support forces were measured by gripping 20%, 40%, and 60% of the stent’s diameter, and bending tests were performed at angles of 60°, 90°, and 180°. The stability of the testing device was analyzed through comparative tests across different compression diameters and bending angles. Results: The device demonstrated high detection precision, stability, and accuracy, with minimal deviation across multiple measurements. The mechanical behavior of the stent observed in both finite element simulations and physical experiments showed consistent results. Conclusions: The testing device developed in this study effectively measures the mechanical changes in large-diameter stent grafts, providing a new reference for testing large-diameter stents.

Authors

Publication Details

Journal
Progress in Medical Devices
Published
2025-09-25
DOI
https://doi.org/10.61189/063815kuibzu
Primary Topic
Aortic aneurysm repair treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design and verification of the testing device for thoracic aortic stent grafts

Zhang Ai-ling, Shiju Yan, Yu Zhou
Progress in Medical Devices
Aortic aneurysm repair treatments
article

Design and verification of the testing device for thoracic aortic stent grafts

Zhang Ai-ling, Shiju Yan, Yu Zhou
article en

Abstract

Objective: To design a testing device for measuring the radial support force and bending spring back force of stent grafts and evaluate its effectiveness. Methods: A radial force and spring-back force testing device was designed to integrate with a tensile testing machine. The radial compression and bending characteristics of stent grafts for thoracic aorta applications were analyzed, and the corresponding conversion formulas were derived. A custom stent ring fixture was fabricated, and a five-wave gradient stent was sewn. Both physical experiments and finite element simulations were conducted. Radial support forces were measured by gripping 20%, 40%, and 60% of the stent’s diameter, and bending tests were performed at angles of 60°, 90°, and 180°. The stability of the testing device was analyzed through comparative tests across different compression diameters and bending angles. Results: The device demonstrated high detection precision, stability, and accuracy, with minimal deviation across multiple measurements. The mechanical behavior of the stent observed in both finite element simulations and physical experiments showed consistent results. Conclusions: The testing device developed in this study effectively measures the mechanical changes in large-diameter stent grafts, providing a new reference for testing large-diameter stents.

Progress in Medical DevicesVol. 3(3)
Openalex Percentile: Top 61%
Aortic aneurysm repair treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.