Objective: Accurate measurement of orofacial muscle strength is essential for phenotyping patients with obstructive sleep apnea (OSA), and particularly those with hypotonic phenotypes. High costs associated with gold standard devices such as the Iowa Oral Performance Instrument (IOPI®) limit their clinical use. This study aims to validate the Sandway® manometer, a low-cost alternative to the IOPI®, by comparing its performance against the IOPI® and the Tongue Digital Spoon (TDS) in both laboratory and clinical settings. Methods: We conducted a two-phase study. In the laboratory phase, pressure readings from the IOPI® and Sandway® devices were compared using standardized force application methods. In the clinical phase, 60 patients with moderate-to-severe untreated OSA underwent orofacial strength evaluation using IOPI®, Sandway®, and TDS devices. Intraclass Correlation Coefficient (ICC), Bland–Altman analysis, and Pearson correlation were used to assess reliability and agreement. Results: The Sandway® showed excellent reliability for anterior tongue and lip strength measures (ICC = 0.978). Bland–Altman plots revealed minimal bias and narrow limits of agreement compared to the IOPI®, indicating strong agreement for both tongue and lip measurements. A high correlation was also observed between Sandway® and TDS results (r = 0.863, p < 0.001), supporting secondary validation. Conclusions: The Sandway® manometer demonstrates high reliability and strong agreement with gold standard instruments, representing a valid, accessible alternative for objective assessment of orofacial muscle strength in clinical practice.

Authors

Institutions

Publication Details

Journal
International Journal of Orofacial Myology
Published
2025-09-29
DOI
https://doi.org/10.3390/ijom51020009
Primary Topic
Obstructive Sleep Apnea Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Objective Assessment of Orofacial Muscle Strength: Validation of an Alternative Low-Cost Measurement Device

Maria Teresa Garcı́a-Iriarte, Eduardo J. Correa, Marta Morato, Guillermo Plaza et al.
International Journal of Orofacial Myology
Obstructive Sleep Apnea Research
article

Objective Assessment of Orofacial Muscle Strength: Validation of an Alternative Low-Cost Measurement Device

Maria Teresa Garcı́a-Iriarte, Eduardo J. Correa, Marta Morato, Guillermo Plaza, James A. Curtis, Carlos O’Connor-Reina, Gislaine Aparecida Folha, Laura Rodríguez Alcalá, Cristina Rodriguez Alcalá, Juan Antonio Ibañez-Rodriguez
article en

Abstract

Objective: Accurate measurement of orofacial muscle strength is essential for phenotyping patients with obstructive sleep apnea (OSA), and particularly those with hypotonic phenotypes. High costs associated with gold standard devices such as the Iowa Oral Performance Instrument (IOPI®) limit their clinical use. This study aims to validate the Sandway® manometer, a low-cost alternative to the IOPI®, by comparing its performance against the IOPI® and the Tongue Digital Spoon (TDS) in both laboratory and clinical settings. Methods: We conducted a two-phase study. In the laboratory phase, pressure readings from the IOPI® and Sandway® devices were compared using standardized force application methods. In the clinical phase, 60 patients with moderate-to-severe untreated OSA underwent orofacial strength evaluation using IOPI®, Sandway®, and TDS devices. Intraclass Correlation Coefficient (ICC), Bland–Altman analysis, and Pearson correlation were used to assess reliability and agreement. Results: The Sandway® showed excellent reliability for anterior tongue and lip strength measures (ICC = 0.978). Bland–Altman plots revealed minimal bias and narrow limits of agreement compared to the IOPI®, indicating strong agreement for both tongue and lip measurements. A high correlation was also observed between Sandway® and TDS results (r = 0.863, p < 0.001), supporting secondary validation. Conclusions: The Sandway® manometer demonstrates high reliability and strong agreement with gold standard instruments, representing a valid, accessible alternative for objective assessment of orofacial muscle strength in clinical practice.

International Journal of Orofacial MyologyVol. 51(2)
Universidade de São Paulo (BR), Cornell University (US), Hospital La Luz (ES), Hospital Universitario Quirónsalud Madrid (ES), Hospital Quirónsalud Barcelona (ES), Hospital Universitario de Valme (ES), Hospital Universitario de Fuenlabrada (ES)
Openalex Percentile: Top 66%
Obstructive Sleep Apnea Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.