Analytical performance and user-friendliness of cobas pulse glucose meter in the hands of the intended users

Diabetes is a growing global health concern affecting millions worldwide. Glucose measurement remains a cornerstone in the diagnosis, monitoring, and management of this disease. Considering that numerous point-of-care (POC) devices for glucose measurements are available, systematic evaluation of their analytical performance and user-friendliness is essential. This study presents the findings of an independent evaluation of the cobas pulse glucose meter (Roche Diagnostics GmbH) conducted by the Scandinavian evaluation of laboratory equipment for point of care testing (SKUP). The study, which assessed both the analytical performance and user-friendliness, included 217 participants with diabetes and was conducted in both a hospital laboratory and primary healthcare settings. Capillary whole blood measurements were compared to a plasma glucose hexokinase method at Vejle hospital, Denmark. Certified reference materials from the National Institute of Standards and Technology (NIST) were used to correct all sample results. Repeatability and accuracy were assessed against predefined analytical performance specifications, and user-friendliness was evaluated using a structured questionnaire developed by SKUP. The cobas pulse demonstrated good repeatability (CV: 1.8-4.0%) and fulfilled the specification for accuracy, with minimal bias (maximum +0.16 mmol/L in the 7-10 mmol/L range). Haematocrit had no significant effect within the tested range (5-70%). User-friendliness was rated satisfactory, attributed to intuitive operation, rapid analysis times, and effective quality control features. In conclusion, the cobas pulse glucose meter met the performance criteria for professional use in clinical and primary care settings, supporting its suitability in near-patient glucose monitoring.

Authors

Institutions

Publication Details

Journal
Scandinavian Journal of Clinical and Laboratory Investigation
Published
2025-08-18
DOI
https://doi.org/10.1080/00365513.2025.2565732
Primary Topic
Diabetes Management and Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Analytical performance and user-friendliness of cobas pulse glucose meter in the hands of the intended users

Heidi Berghäll, Dår Kristian Kur, Anne Stavelin, Kristiina Kurg et al.
Scandinavian Journal of Clinical and Laboratory Investigation
Diabetes Management and Research
article

Analytical performance and user-friendliness of cobas pulse glucose meter in the hands of the intended users

Heidi Berghäll, Dår Kristian Kur, Anne Stavelin, Kristiina Kurg, Ričardas Stonys, Johan Olausson, Anne Elisabeth Solsvik, Johan Baden Birk‐Korch, Kirsi A. Virtanen, Anne Haapala, Cindhya Sithiravel, G. Fosse, Sander Johannes Thorbjørnsen Guttorm, Christine Morken, Gustav Hallén, Joakim Hekland, Tine-Lise Kalleklev, Stiina Kamula, Elisabeth Wiik Vigerust, Johannes Østrem Fjøse
article en

Abstract

Diabetes is a growing global health concern affecting millions worldwide. Glucose measurement remains a cornerstone in the diagnosis, monitoring, and management of this disease. Considering that numerous point-of-care (POC) devices for glucose measurements are available, systematic evaluation of their analytical performance and user-friendliness is essential. This study presents the findings of an independent evaluation of the cobas pulse glucose meter (Roche Diagnostics GmbH) conducted by the Scandinavian evaluation of laboratory equipment for point of care testing (SKUP). The study, which assessed both the analytical performance and user-friendliness, included 217 participants with diabetes and was conducted in both a hospital laboratory and primary healthcare settings. Capillary whole blood measurements were compared to a plasma glucose hexokinase method at Vejle hospital, Denmark. Certified reference materials from the National Institute of Standards and Technology (NIST) were used to correct all sample results. Repeatability and accuracy were assessed against predefined analytical performance specifications, and user-friendliness was evaluated using a structured questionnaire developed by SKUP. The cobas pulse demonstrated good repeatability (CV: 1.8-4.0%) and fulfilled the specification for accuracy, with minimal bias (maximum +0.16 mmol/L in the 7-10 mmol/L range). Haematocrit had no significant effect within the tested range (5-70%). User-friendliness was rated satisfactory, attributed to intuitive operation, rapid analysis times, and effective quality control features. In conclusion, the cobas pulse glucose meter met the performance criteria for professional use in clinical and primary care settings, supporting its suitability in near-patient glucose monitoring.

Scandinavian Journal of Clinical and Laboratory InvestigationVol. 85(6)
Oslo University Hospital (NO), University Hospital of North Norway (NO), Karolinska University Hospital (SE), Vilnius University (LT), OsloMet – Oslo Metropolitan University (NO), Akershus University Hospital (NO), Odense University Hospital (DK), Deaconess Hospital (US), Östersunds Hospital (SE), NordLab (FI), Fimlab (Finland) (FI), National Institute of Quality (PE), Lillebaelt Hospital (DK), Euroquality (FR), Vilnius University Hospital Santariskiu Klinikos (LT), Universidad Metropolitana (PR), University of Tartu (EE), UiT The Arctic University of Norway (NO)
Openalex Percentile: Top 64%
Diabetes Management and Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.