Ultrafine Molybdenum Wire Braided Neurointerventional Implants: Bridging Biodegradability and Neurovascular Safety for Stroke Treatment

Neurovascular implants for stroke intervention face a critical dilemma: permanent devices (e.g., nitinol stents, platinum coils) often trigger chronic inflammation and recurrence, whereas biodegradable alternatives (Mg, Fe, Zn alloys) lack radiopacity or raise neurotoxicity concerns. Here, we introduce φ50 µm molybdenum (Mo) wire braided implants that integrate procedural efficacy with biological safety. Mo demonstrates negligible hemolysis (<5%), platelet-inert surfaces, and preserved coagulation kinetics, together with robust cytocompatibility across neurovascular unit cells (endothelia, astrocytes, neurons) under both physiological and ischemia-reperfusion conditions. In vivo, Mo stent wires implanted in rodent carotids maintained blood homeostasis, organ integrity, and neurological function without systemic toxicity. Moreover, braided 2D Mo coils achieved durable aneurysm occlusion with controlled inflammatory resolution and progressive endothelialization, closely resembling clinical performance. Importantly, Mo ions showed no detectable accumulation in brain, kidney, lung, or spleen, attributable to renal clearance and blood-brain barrier selectivity. By coupling intrinsic radiopacity with homogeneous, moderate corrosion, Mo addresses long-standing limitations of existing biodegradable alloys. These findings position Mo as a transformative candidate for next-generation neurovascular devices, harmonizing biodegradability, safety, and imaging precision to redefine the management of both ischemic and hemorrhagic stroke.

Authors

Institutions

Publication Details

Journal
Advanced Science
Published
2025-09-29
DOI
https://doi.org/10.1002/advs.202511466
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Ultrafine Molybdenum Wire Braided Neurointerventional Implants: Bridging Biodegradability and Neurovascular Safety for Stroke Treatment

Wu Wang, Dong Bian, Di Wu, Y. Zhang et al.
Advanced Science
Orthopaedic implants and arthroplasty
article

Ultrafine Molybdenum Wire Braided Neurointerventional Implants: Bridging Biodegradability and Neurovascular Safety for Stroke Treatment

Wu Wang, Dong Bian, Di Wu, Y. Zhang, Yiming Huang, Yufeng Zheng, Yunong Shen, Chunhao Yu, Heng Liu, Li Ming, Xunming Ji, Miaowen Jiang
article en

Abstract

Neurovascular implants for stroke intervention face a critical dilemma: permanent devices (e.g., nitinol stents, platinum coils) often trigger chronic inflammation and recurrence, whereas biodegradable alternatives (Mg, Fe, Zn alloys) lack radiopacity or raise neurotoxicity concerns. Here, we introduce φ50 µm molybdenum (Mo) wire braided implants that integrate procedural efficacy with biological safety. Mo demonstrates negligible hemolysis (<5%), platelet-inert surfaces, and preserved coagulation kinetics, together with robust cytocompatibility across neurovascular unit cells (endothelia, astrocytes, neurons) under both physiological and ischemia-reperfusion conditions. In vivo, Mo stent wires implanted in rodent carotids maintained blood homeostasis, organ integrity, and neurological function without systemic toxicity. Moreover, braided 2D Mo coils achieved durable aneurysm occlusion with controlled inflammatory resolution and progressive endothelialization, closely resembling clinical performance. Importantly, Mo ions showed no detectable accumulation in brain, kidney, lung, or spleen, attributable to renal clearance and blood-brain barrier selectivity. By coupling intrinsic radiopacity with homogeneous, moderate corrosion, Mo addresses long-standing limitations of existing biodegradable alloys. These findings position Mo as a transformative candidate for next-generation neurovascular devices, harmonizing biodegradability, safety, and imaging precision to redefine the management of both ischemic and hemorrhagic stroke.

Advanced ScienceVol. 13(1)
Beijing Institute of Technology (CN), Capital Medical University (CN), Peking University (CN), Shanghai University of Traditional Chinese Medicine (CN), Longhua Hospital Shanghai University of Traditional Chinese Medicine (CN), Chinese Institute for Brain Research (CN), Guangdong Academy of Medical Sciences (CN), Guangdong Provincial People's Hospital (CN), Southern Medical University (CN)
Openalex Percentile: Top 57%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.