Developing Low-Viscosity Lubricants to Maintain Lubricity between Overtube and Enteroscope for Improved Maneuverability during Single-Balloon Enteroscopy

During enteroscopy, the hydrophilic coating on the inner surface of the overtube (OT) reduces the dynamic friction coefficient (DFC) between the endoscopic shaft (ES) and OT, allowing smooth insertion of the ES. However, endoscopists often encounter resistance to ES insertion and poor endoscopic manipulation during prolonged enteroscopy. We hypothesized that the reason is the degradation of the hydrophilic coating due to the continuous contact between the ES and OT and the associated increase in DFC. Focusing on single-balloon enteroscopy, we developed a unique model to continuously measure the DFC for 30 min while moving the ES in contact with the inner surface of the OT, which was applied with pure water or various lubricants. When pure water was applied, the initial DFC was 0.112 ± 0.001; however, it increased to 0.357 ± 0.002 over 30 min (219% increase). Meanwhile, the application of 0.1% carboxymethyl cellulose, 0.05% sodium alginate, 0.25% hydroxyethyl cellulose, and 0.04% xanthan gum solutions resulted in significantly lower initial DFCs of 0.073 ± 0.0004, 0.081 ± 0.0004, 0.079 ± 0.0004, and 0.063 ± 0.0003, respectively. Moreover, low DFCs were maintained even after 30 min, with DFC increase rates of +2.5%, -0.2%, -5.9%, and +5.6%, respectively. In addition, the injection pressures required for these optimal lubricants were extremely low. The optimal lubricants identified in this study successfully maintained lower DFCs than pure water, which is conventionally used to induce lubrication between the ES and OT; therefore, they may improve endoscopic manipulation.

Authors

Institutions

Publication Details

Journal
ACS Omega
Published
2025-09-30
DOI
https://doi.org/10.1021/acsomega.5c06942
Primary Topic
Gastrointestinal Bleeding Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Developing Low-Viscosity Lubricants to Maintain Lubricity between Overtube and Enteroscope for Improved Maneuverability during Single-Balloon Enteroscopy

Katsuma Yamauchi, Takaaki Nakaya, Ryohei Hirose, Naohisa Yoshida et al.
ACS Omega
Gastrointestinal Bleeding Diagnosis and Treatment
article

Developing Low-Viscosity Lubricants to Maintain Lubricity between Overtube and Enteroscope for Improved Maneuverability during Single-Balloon Enteroscopy

Katsuma Yamauchi, Takaaki Nakaya, Ryohei Hirose, Naohisa Yoshida, Ken Inoue, Takehiro Kanou, Akinobu Sai, Osamu Dohi, Naoto Watanabe, Yoshito Itoh
article en

Abstract

During enteroscopy, the hydrophilic coating on the inner surface of the overtube (OT) reduces the dynamic friction coefficient (DFC) between the endoscopic shaft (ES) and OT, allowing smooth insertion of the ES. However, endoscopists often encounter resistance to ES insertion and poor endoscopic manipulation during prolonged enteroscopy. We hypothesized that the reason is the degradation of the hydrophilic coating due to the continuous contact between the ES and OT and the associated increase in DFC. Focusing on single-balloon enteroscopy, we developed a unique model to continuously measure the DFC for 30 min while moving the ES in contact with the inner surface of the OT, which was applied with pure water or various lubricants. When pure water was applied, the initial DFC was 0.112 ± 0.001; however, it increased to 0.357 ± 0.002 over 30 min (219% increase). Meanwhile, the application of 0.1% carboxymethyl cellulose, 0.05% sodium alginate, 0.25% hydroxyethyl cellulose, and 0.04% xanthan gum solutions resulted in significantly lower initial DFCs of 0.073 ± 0.0004, 0.081 ± 0.0004, 0.079 ± 0.0004, and 0.063 ± 0.0003, respectively. Moreover, low DFCs were maintained even after 30 min, with DFC increase rates of +2.5%, -0.2%, -5.9%, and +5.6%, respectively. In addition, the injection pressures required for these optimal lubricants were extremely low. The optimal lubricants identified in this study successfully maintained lower DFCs than pure water, which is conventionally used to induce lubrication between the ES and OT; therefore, they may improve endoscopic manipulation.

ACS OmegaVol. 10(40)
Kyoto Prefectural University of Medicine (JP)
Openalex Percentile: Top 61%
Gastrointestinal Bleeding Diagnosis and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.