Impact of Thickness and Composition on YAG:Ce Composite Scintillators for Alpha Particle Detectors
This study evaluates the performance of imaging-based alpha particle detectors using YAG:Ce composite scintillators in mixed radiation fields. Scintillators with varying thicknesses (0.1 mm, 0.5 mm, and 1.0 mm) and YAG:Ce to epoxy ratios (5%, 15%, and 20%) were modeled using the GEANT4 simulation package to analyze the detection efficiency, light output, and energy resolution of the alpha particle detectors. The results indicate that a 0.5 mm thick scintillator with 90% YAG:Ce and 10% epoxy is the optimal configuration among the simulated scintillators for an alpha particle detector. Additionally, phantom-based simulations demonstrated that the detector is capable of producing clear imaging of alpha particles.
Authors
- O. B. Kolcu (ORCID: https://orcid.org/0000-0002-9177-1286)
Institutions
- Istinye University (TR)
Publication Details
- Journal
- Celal Bayar Üniversitesi Fen Bilimleri Dergisi
- Published
- 2025-09-26
- DOI
- https://doi.org/10.18466/cbayarfbe.1599240
- Primary Topic
- Radiation Detection and Scintillator Technologies
- Type
- article
- Field-Weighted Citation Impact
- 0.00