Evaluation of cutting edge integrity, surface degradation, and material loss in four implant drilling systems: an in vitro investigation in bovine rib bone

Implant drills are essential for precise osteotomy preparation, and their repeated use can affect surface integrity, sharpness, and material loss. This original study presents a comprehensive evaluation of four implant drill systems—conventional stainless steel drills (CS), hard-coated stainless steel drills (HCS), ceramic drills (CD), and tapered expansion drills (TE)—by assessing cutting edge wear, surface roughness, and material loss at defined usage intervals. A total of 160 osteotomies were performed on fresh bovine rib bone blocks using four different drill systems. Each group included 40 repetitions. Each drill was analyzed at baseline and after 10, 20, 30, and 40 osteotomies. Drilling was performed under standardized parameters (23 °C, 800 rpm, 2 kg axial load, 50 mL/min irrigation). Surface roughness was measured using a stylus-based contact profilometer. SEM was used for morphological inspection, and a precision analytical balance (± 0.0001 g) was used to detect material loss. No significant material loss was observed in any group during the first 20 osteotomies. The ceramic drills demonstrated zero material loss and minimal surface wear across all time points. HCS drills showed the highest variability in mass loss (up to 136.5 µg), and TE drills exhibited visible edge degradation after 30 uses. SEM revealed progressive morphological changes, including edge rounding and microcracks, predominantly in HCS and TE groups. While differences between groups were not statistically significant, ceramic drills consistently demonstrated superior structural preservation. Repeated use of implant drills results in gradual wear, particularly in metallic systems. Ceramic drills exhibited the most stable performance, suggesting their potential advantage for extended clinical use.

Authors

Institutions

Publication Details

Journal
BMC Oral Health
Published
2025-09-29
DOI
https://doi.org/10.1186/s12903-025-06896-1
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of cutting edge integrity, surface degradation, and material loss in four implant drilling systems: an in vitro investigation in bovine rib bone

Mehmet Ali Erdem, Başak Keskin Yalçın, Betül Gedik, Hülya Koçak Berberoğlu et al.
BMC Oral Health
Dental Implant Techniques and Outcomes
article

Evaluation of cutting edge integrity, surface degradation, and material loss in four implant drilling systems: an in vitro investigation in bovine rib bone

Mehmet Ali Erdem, Başak Keskin Yalçın, Betül Gedik, Hülya Koçak Berberoğlu, Abdulkadir Burak Cankaya, Eda Etik
article en

Abstract

Implant drills are essential for precise osteotomy preparation, and their repeated use can affect surface integrity, sharpness, and material loss. This original study presents a comprehensive evaluation of four implant drill systems—conventional stainless steel drills (CS), hard-coated stainless steel drills (HCS), ceramic drills (CD), and tapered expansion drills (TE)—by assessing cutting edge wear, surface roughness, and material loss at defined usage intervals. A total of 160 osteotomies were performed on fresh bovine rib bone blocks using four different drill systems. Each group included 40 repetitions. Each drill was analyzed at baseline and after 10, 20, 30, and 40 osteotomies. Drilling was performed under standardized parameters (23 °C, 800 rpm, 2 kg axial load, 50 mL/min irrigation). Surface roughness was measured using a stylus-based contact profilometer. SEM was used for morphological inspection, and a precision analytical balance (± 0.0001 g) was used to detect material loss. No significant material loss was observed in any group during the first 20 osteotomies. The ceramic drills demonstrated zero material loss and minimal surface wear across all time points. HCS drills showed the highest variability in mass loss (up to 136.5 µg), and TE drills exhibited visible edge degradation after 30 uses. SEM revealed progressive morphological changes, including edge rounding and microcracks, predominantly in HCS and TE groups. While differences between groups were not statistically significant, ceramic drills consistently demonstrated superior structural preservation. Repeated use of implant drills results in gradual wear, particularly in metallic systems. Ceramic drills exhibited the most stable performance, suggesting their potential advantage for extended clinical use.

BMC Oral HealthVol. 25(1)
Istanbul University (TR)
Openalex Percentile: Top 63%
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.