An Engineered Au-MXene Sorption Trap Enables Dual-Channel Ultrafast and Selective Detection of Trace Gaseous Elemental Mercury
Accurate detection of trace atmospheric mercury is critical for toxicological assessment and environmental protection. However, achieving parts per billion-level sensitivity with high selectivity between gaseous elemental mercury (Hg<sup>0</sup>) and oxidized mercury (Hg<sup>2+</sup>) remains a significant challenge. Here, we present a novel dual-channel enrichment trap system that selectively captures Hg<sup>0</sup> by using a layered Au-MXene (Ti<sub>3</sub>C<sub>2</sub>T<sub><i>x</i></sub>) composite as the sorption trap. This composite demonstrates exceptional selectivity for Hg<sup>0</sup>, with a selectivity ratio of up to 76.7 over Hg<sup>2+</sup>, and achieves 99.7% ± 0.3% (<i>n</i> = 3) sorption efficiency for ppb-level Hg<sup>0</sup> within 30 min at room temperature. Complete desorption is achieved in just 90 s at 300 °C, demonstrating a response time of 45 s for Hg<sup>0</sup> detection through the dual-electrothermal Hg<sup>0</sup> channel. The Au-Ti<sub>3</sub>C<sub>2</sub>T<sub><i>x</i></sub> preconcentration cartridge exhibits excellent moisture resistance, high-temperature stability, and long-term durability. Density functional theory (DFT) calculations reveal that the built-in electric field at the Au-Ti<sub>3</sub>C<sub>2</sub>T<sub><i>x</i></sub> interface accelerates electron transfer, reducing preconcentration and regeneration times. This technology integrates seamlessly with the cold vapor atomic absorption spectrophotometer (CVAAS) method to form a mercury dual-channel continuous emission measurement system (Hg DCEMS), ensuring highly selective and precise trace Hg<sup>0</sup> detection. This advancement provides a device for accurately measuring the Hg<sup>0</sup> concentration, enabling better mercury pollution control.
Authors
- Zan Qu (ORCID: https://orcid.org/0000-0003-2193-2715)
- Naiqiang Yan (ORCID: https://orcid.org/0000-0002-7821-1997)
- Qingyan Fu (ORCID: https://orcid.org/0000-0003-2192-5654)
- Wenjun Huang (ORCID: https://orcid.org/0000-0001-8887-309X)
- Qianxuan Zhang
- Jia Chen
- Yiqi Feng
- Haomiao Xu (ORCID: https://orcid.org/0000-0003-3736-4548)
- Bing Pei (ORCID: https://orcid.org/0000-0003-4435-8701)
- Qinyuan Hong (ORCID: https://orcid.org/0000-0002-0733-7402)
- Menggai Jiao (ORCID: https://orcid.org/0009-0006-6691-1837)
- Li Yang (ORCID: https://orcid.org/0000-0001-6569-2886)
Institutions
- Shanghai Jiao Tong University (CN)
- Ministry of Ecology and Environment (CN)
- Beijing Founder Electronics (China) (CN)
- Zhejiang Environmental Monitoring Center (CN)
- Shanghai Academy of Environmental Sciences (CN)
- Shanghai Institute of Pollution Control and Ecological Security
Publication Details
- Journal
- Analytical Chemistry
- Published
- 2025-09-27
- DOI
- https://doi.org/10.1021/acs.analchem.5c02853
- Primary Topic
- MXene and MAX Phase Materials
- Type
- article
- Field-Weighted Citation Impact
- 0.00