Surgical repair for a single leaflet device attachment due to a gripper failure from the MitraClip main frame: a case report
To our knowledge, gripper detachment has not previously been reported as a cause of single leaflet device attachment. While echocardiography may be limited in detecting device-related failures preoperatively, computed tomography may aid in identifying retained device fragments before the procedure. This case highlights the importance of understanding device structure and anticipating rare mechanical failures when planning reintervention following MitraClip therapy.
Authors
- Daisuke Futagami
- Hironobu Morimoto
- Shogo Mukai
- Takeshi Harada
- Keijiro Katayama
Institutions
- Fukuyama City Hospital (JP)
- Fukuyama Cardiovascular Hospital (JP)
Publication Details
- Journal
- European Heart Journal - Case Reports
- Published
- 2025-09-27
- DOI
- https://doi.org/10.1093/ehjcr/ytaf489
- Primary Topic
- Cardiac Valve Diseases and Treatments
- Type
- article
- Field-Weighted Citation Impact
- 0.00