Influence of radial clearance on Tresca stress in Al2O2-on-Al2O3 bearings for total hip prosthesis evaluated using finite element analysis

Ceramic materials have gained prominence as advanced alternatives to metals and polymers in total hip prosthesis, owing to their superior wear resistance, chemical inertness, and minimal ion release. Among ceramic biomaterials, aluminum oxide (Al<sub>2</sub>O<sub>2</sub>) is particularly favored for its exceptional hardness, high compressive strength, and excellent tribological performance, outperforming other candidates such as zirconium dioxide (ZrO₂) and silicon nitride (Si<sub>2</sub>N<sub>4</sub>). However, the inherently brittle nature of ceramics necessitates careful stress analysis to mitigate fracture risks in ceramic-on-ceramic bearing systems. This study investigates the influence of radial clearance as a key geometric parameter on the mechanical safety of Al<sub>2</sub>O<sub>2</sub>-on-Al<sub>2</sub>O<sub>3</sub> bearings by analyzing Tresca stress distributions under physiologically representative loading conditions. A 2D axisymmetric finite element model was developed to simulate the internal stress behavior of the bearing components during a full gait cycle. Six radial clearance values, ranging from 0.03 to 0.3 mm, were systematically evaluated to determine their effect on internal shear stress patterns. The simulation results demonstrate that a radial clearance of 0.03 mm yields the lowest Tresca stress, indicating a more favorable stress distribution and reduced risk of shear-induced fracture. Conversely, larger clearances were associated with elevated shear stresses, which could compromise the mechanical integrity of the ceramic components over time. These findings highlight the importance of precise control over radial clearance during the design and manufacturing of ceramic-on-ceramic hip implants. The study provides foundational insights for optimizing implant geometry to enhance the structural reliability and clinical longevity of ceramic-bearing total hip prostheses.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-09-26
DOI
https://doi.org/10.1038/s41598-025-16850-4
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of radial clearance on Tresca stress in Al2O2-on-Al2O3 bearings for total hip prosthesis evaluated using finite element analysis

Mohamad Izzur Maula, Muhammad Hanif Ramlee, Muhammad Kozin, Abdulfatah Abdu Yusuf et al.
Scientific Reports
Orthopaedic implants and arthroplasty
article

Influence of radial clearance on Tresca stress in Al2O2-on-Al2O3 bearings for total hip prosthesis evaluated using finite element analysis

Mohamad Izzur Maula, Muhammad Hanif Ramlee, Muhammad Kozin, Abdulfatah Abdu Yusuf, J. Jamari, M. Danny Pratama Lamura, Athanasius Priharyoto Bayuseno, Muhammad Imam Ammarullah, Hasyid Ahmad Wicaksono
article en

Abstract

Ceramic materials have gained prominence as advanced alternatives to metals and polymers in total hip prosthesis, owing to their superior wear resistance, chemical inertness, and minimal ion release. Among ceramic biomaterials, aluminum oxide (Al<sub>2</sub>O<sub>2</sub>) is particularly favored for its exceptional hardness, high compressive strength, and excellent tribological performance, outperforming other candidates such as zirconium dioxide (ZrO₂) and silicon nitride (Si<sub>2</sub>N<sub>4</sub>). However, the inherently brittle nature of ceramics necessitates careful stress analysis to mitigate fracture risks in ceramic-on-ceramic bearing systems. This study investigates the influence of radial clearance as a key geometric parameter on the mechanical safety of Al<sub>2</sub>O<sub>2</sub>-on-Al<sub>2</sub>O<sub>3</sub> bearings by analyzing Tresca stress distributions under physiologically representative loading conditions. A 2D axisymmetric finite element model was developed to simulate the internal stress behavior of the bearing components during a full gait cycle. Six radial clearance values, ranging from 0.03 to 0.3 mm, were systematically evaluated to determine their effect on internal shear stress patterns. The simulation results demonstrate that a radial clearance of 0.03 mm yields the lowest Tresca stress, indicating a more favorable stress distribution and reduced risk of shear-induced fracture. Conversely, larger clearances were associated with elevated shear stresses, which could compromise the mechanical integrity of the ceramic components over time. These findings highlight the importance of precise control over radial clearance during the design and manufacturing of ceramic-on-ceramic hip implants. The study provides foundational insights for optimizing implant geometry to enhance the structural reliability and clinical longevity of ceramic-bearing total hip prostheses.

Scientific ReportsVol. 15(1)
Diponegoro University (ID), University of Liberia (LR), University of Pembangunan Nasional Veteran Jawa Timur (ID), National Research and Innovation Agency (ID), University of Technology Malaysia (MY)
Universitas Diponegoro
Openalex Percentile: Top 58%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.