Metal artifact reduction and contrast agent performance as a function of source spectral shape

<i>Objective.</i>Empirically map the x-ray source of a Medipix All Resolution Scanner (MARS) photon counting CT (PCCT) with Cu and Sn pre-filters, assess metal artifact reduction (MAR) capabilities of these pre-filters, and measure pre-filtration impact on contrast performance (CP) of FDA approved iopromide and experimental tantalum oxide nanoparticles (TaO<i><sub>x</sub></i>NPs).<i>Approach.</i>The x-ray source of a MARS-PCCT system was empirically mapped with no pre-filtration, seven Cu filters (0.3-2.1 mm), and seven Sn filters (0.15-1.05 mm). A phantom with inserts containing water, lipid, iopromide, TaO<i><sub>x</sub></i>NPs, and metal was scanned with no pre-filtration and Cu and Sn pre-filters. Insert noise, signal, contrast-to-pooled-noise ratio (CPNR), and a fast Fourier transform artifact metric (FFTAM) were calculated for each filter to quantify MAR and CP.<i>Main results.</i>Thick filters for Cu and Sn shifted mean energy of the unfiltered x-ray source (47.9 keV) by 19.2 and 23.4 keV, respectively. Thick filtration, 2.1 mm Cu and 0.6 mm Sn, greatly reduced noise (up to 74%) and FFTAM (up to 71%) for all inserts and energy bins. Thin filtration, 0.3 mm Cu and 0.15 mm Sn, also reduced noise (up to 47%) and FFTAM (up to 41%). In most cases, iopromide lost significant contrast (up to 50%). TaO<i><sub>x</sub></i>NPs also lost contrast, though to a lesser extent (up to 38%). Pre-filtration improved image efficacy (i.e. CPNR), especially for TaO<i><sub>x</sub></i>(up to 61%).<i>Significance.</i>By empirically mapping the source spectrum of a MARS-PCCT system with pre-filters, valuable information was gathered about photon flux distribution and detector artifacts; these findings will prove insightful for applications such as energy binning for effective material decomposition. Furthermore, this information will potentially guide clinical MAR development, most notably for the MARS Extremity 5 × 120 recently deployed for first-in-human trials. Lastly, TaO<i><sub>x</sub></i>NPs were shown to be more compatible than iopromide with spectral shaping.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-09-25
DOI
https://doi.org/10.1088/1361-6560/ae0be9
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Metal artifact reduction and contrast agent performance as a function of source spectral shape

Peter J. Bonitatibus, Ge Wang, Jacob Daniel Aubrey, Emmett Patrick Perkinson
Physics in Medicine and Biology
Advanced X-ray and CT Imaging
article

Metal artifact reduction and contrast agent performance as a function of source spectral shape

Peter J. Bonitatibus, Ge Wang, Jacob Daniel Aubrey, Emmett Patrick Perkinson
article en

Abstract

<i>Objective.</i>Empirically map the x-ray source of a Medipix All Resolution Scanner (MARS) photon counting CT (PCCT) with Cu and Sn pre-filters, assess metal artifact reduction (MAR) capabilities of these pre-filters, and measure pre-filtration impact on contrast performance (CP) of FDA approved iopromide and experimental tantalum oxide nanoparticles (TaO<i><sub>x</sub></i>NPs).<i>Approach.</i>The x-ray source of a MARS-PCCT system was empirically mapped with no pre-filtration, seven Cu filters (0.3-2.1 mm), and seven Sn filters (0.15-1.05 mm). A phantom with inserts containing water, lipid, iopromide, TaO<i><sub>x</sub></i>NPs, and metal was scanned with no pre-filtration and Cu and Sn pre-filters. Insert noise, signal, contrast-to-pooled-noise ratio (CPNR), and a fast Fourier transform artifact metric (FFTAM) were calculated for each filter to quantify MAR and CP.<i>Main results.</i>Thick filters for Cu and Sn shifted mean energy of the unfiltered x-ray source (47.9 keV) by 19.2 and 23.4 keV, respectively. Thick filtration, 2.1 mm Cu and 0.6 mm Sn, greatly reduced noise (up to 74%) and FFTAM (up to 71%) for all inserts and energy bins. Thin filtration, 0.3 mm Cu and 0.15 mm Sn, also reduced noise (up to 47%) and FFTAM (up to 41%). In most cases, iopromide lost significant contrast (up to 50%). TaO<i><sub>x</sub></i>NPs also lost contrast, though to a lesser extent (up to 38%). Pre-filtration improved image efficacy (i.e. CPNR), especially for TaO<i><sub>x</sub></i>(up to 61%).<i>Significance.</i>By empirically mapping the source spectrum of a MARS-PCCT system with pre-filters, valuable information was gathered about photon flux distribution and detector artifacts; these findings will prove insightful for applications such as energy binning for effective material decomposition. Furthermore, this information will potentially guide clinical MAR development, most notably for the MARS Extremity 5 × 120 recently deployed for first-in-human trials. Lastly, TaO<i><sub>x</sub></i>NPs were shown to be more compatible than iopromide with spectral shaping.

Physics in Medicine and BiologyVol. 70(20)
Rensselaer Polytechnic Institute (US), Hudson Valley Community College (US)
Openalex Percentile: Top 75%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.