Evaluation of Diagnostic Performance of the Transferrin/FOB Combo Rapid Test Cassette (Feces) for Gastrointestinal Bleeding

The Transferrin/FOB Combo Rapid Test Cassette (Feces) is a chromatographic immunoassay designed for the qualitative detection of human hemoglobin and transferrin in human fecal specimens. This study aims to evaluate the consistency between test results of Transferrin/FOB Combo rapid test produced by CITEST DIAGNOSTCS INC. and the contrast test device, and evaluate its clinical application ability through testing on certain quantity of representative specimens and through scientific and reasonable statistic analysis on the results. A total of 436 fecal specimens from both symptomatic and asymptomatic individuals undergoing endoscopic examination were tested. The reference method was a commercially available rapid test (W.H.P.M Inc. Transferrin/Hemoglobin FOB Combo Rapid Test). For FOB detection, the relative sensitivity was 97.9% (95%CI: 94.1%~99.6%), relative specificity was 99.7% (95%CI: 98.1%~99.9%), and accuracy was 99.1% (95%CI: 97.7%~99.8%), with a Kappa value of 0.98 indicating excellent consistency. For transferrin detection, the relative sensitivity was 98.9% (95%CI: 94.1%~99.9%), relative specificity was 99.4% (95%CI: 97.9%~99.9%) and accuracy was 99.3% (95%CI: 98.0%~99.9%), with a Kappa value of 0.98. The LOD was 50 ng/mL or 6 μg/g feces for FOB and 40 ng/mL or 4 μg/g feces for transferrin. Intra-assay and inter-assay precision were excellent (>99% correct identification). No cross-reactivity was observed with common interfering substances. These findings indicate that the Transferrin/FOB Combo Rapid Test Cassette is a reliable, rapid and practical tool for simultaneous detection of FOB and transferrin, aiding in the timely diagnosis of gastrointestinal bleeding, including upper and lower gastrointestinal disorders.

Publication Details

Journal
MEDS Clinical Medicine
Published
2025-01-01
DOI
https://doi.org/10.23977/medsc.2025.060420
Primary Topic
Gastrointestinal Bleeding Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of Diagnostic Performance of the Transferrin/FOB Combo Rapid Test Cassette (Feces) for Gastrointestinal Bleeding

MEDS Clinical Medicine
Gastrointestinal Bleeding Diagnosis and Treatment
article

Evaluation of Diagnostic Performance of the Transferrin/FOB Combo Rapid Test Cassette (Feces) for Gastrointestinal Bleeding

article en

Abstract

The Transferrin/FOB Combo Rapid Test Cassette (Feces) is a chromatographic immunoassay designed for the qualitative detection of human hemoglobin and transferrin in human fecal specimens. This study aims to evaluate the consistency between test results of Transferrin/FOB Combo rapid test produced by CITEST DIAGNOSTCS INC. and the contrast test device, and evaluate its clinical application ability through testing on certain quantity of representative specimens and through scientific and reasonable statistic analysis on the results. A total of 436 fecal specimens from both symptomatic and asymptomatic individuals undergoing endoscopic examination were tested. The reference method was a commercially available rapid test (W.H.P.M Inc. Transferrin/Hemoglobin FOB Combo Rapid Test). For FOB detection, the relative sensitivity was 97.9% (95%CI: 94.1%~99.6%), relative specificity was 99.7% (95%CI: 98.1%~99.9%), and accuracy was 99.1% (95%CI: 97.7%~99.8%), with a Kappa value of 0.98 indicating excellent consistency. For transferrin detection, the relative sensitivity was 98.9% (95%CI: 94.1%~99.9%), relative specificity was 99.4% (95%CI: 97.9%~99.9%) and accuracy was 99.3% (95%CI: 98.0%~99.9%), with a Kappa value of 0.98. The LOD was 50 ng/mL or 6 μg/g feces for FOB and 40 ng/mL or 4 μg/g feces for transferrin. Intra-assay and inter-assay precision were excellent (>99% correct identification). No cross-reactivity was observed with common interfering substances. These findings indicate that the Transferrin/FOB Combo Rapid Test Cassette is a reliable, rapid and practical tool for simultaneous detection of FOB and transferrin, aiding in the timely diagnosis of gastrointestinal bleeding, including upper and lower gastrointestinal disorders.

MEDS Clinical MedicineVol. 6(4)
Openalex Percentile: Top 62%
Gastrointestinal Bleeding Diagnosis and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.