Effect of Varying HVL Values on the Dose Output of Plain X-Ray Machines at a Fixed kV of 80, a Regulatory Perspective

Background: HVL in diagnostic X-ray machines is an important property for defining the penetration ability of an X-ray beam into the human body and, hence, the beam hardness. Materials and Methods: The study utilized the findings of the radiation safety inspections conducted between 2022 and 2023 to investigate the variation of the Half-value layer (HVL) with dose output, taking 64 X-ray machines based on a set criterion. Results: A total of 26 X-ray machines failed the HVL test, while 38 passed the HVL test. Of the 38 that passed the test, 14 had very high and extremely high HVLs and produced doses below the lower limit of the permissible range of 0.025 - 0.08 mGy/mAs. The corrective action for 26 machines that generally failed the HVL test was to minimize the filtration by removing the filter plates. However, this had to be done only after a conclusive investigation to check the accuracy of tube current (mA), timer (s), and tube potential (kVp) parameters. Conclusion: Therefore, the regulatory body recommends that HVL tests be part of the acceptance and commissioning tests for the new machines in Uganda and be done routinely for the machines in use, as specified in the quality control program for each X-ray machine.

Authors

Institutions

Publication Details

Journal
Open Journal of Medical Imaging
Published
2025-01-01
DOI
https://doi.org/10.4236/ojmi.2025.153017
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of Varying HVL Values on the Dose Output of Plain X-Ray Machines at a Fixed kV of 80, a Regulatory Perspective

Wellen Rukundo, Natharius Nimbashabira, Noah Deogratias Luwalira
Open Journal of Medical Imaging
Advanced X-ray and CT Imaging
article

Effect of Varying HVL Values on the Dose Output of Plain X-Ray Machines at a Fixed kV of 80, a Regulatory Perspective

Wellen Rukundo, Natharius Nimbashabira, Noah Deogratias Luwalira
article en

Abstract

Background: HVL in diagnostic X-ray machines is an important property for defining the penetration ability of an X-ray beam into the human body and, hence, the beam hardness. Materials and Methods: The study utilized the findings of the radiation safety inspections conducted between 2022 and 2023 to investigate the variation of the Half-value layer (HVL) with dose output, taking 64 X-ray machines based on a set criterion. Results: A total of 26 X-ray machines failed the HVL test, while 38 passed the HVL test. Of the 38 that passed the test, 14 had very high and extremely high HVLs and produced doses below the lower limit of the permissible range of 0.025 - 0.08 mGy/mAs. The corrective action for 26 machines that generally failed the HVL test was to minimize the filtration by removing the filter plates. However, this had to be done only after a conclusive investigation to check the accuracy of tube current (mA), timer (s), and tube potential (kVp) parameters. Conclusion: Therefore, the regulatory body recommends that HVL tests be part of the acceptance and commissioning tests for the new machines in Uganda and be done routinely for the machines in use, as specified in the quality control program for each X-ray machine.

Open Journal of Medical ImagingVol. 15(03)
Atomic Energy Council (TW)
Openalex Percentile: Top 75%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Effect of Varying HVL Values on the Dose Output of Plain X-Ray Machines at a Fixed kV of 80, a Regulatory Perspective — Wellen Rukundo, Natharius Nimbashabira, et al. · Open Journal of Medical Imaging (2025) | TGRS Research Map | TGRS