Body Size Modulates the Impact of the Dispersive Patch Position During Radiofrequency Cardiac Ablation

(1) Background: In the context of cardiac radiofrequency (RF) ablation, it has been proposed that positioning the dispersive patch (DP) concordantly with the orientation of the ablation electrode may enhance lesion size. The objective of this study is to investigate how individual body size may modulate the extent of this effect. (2) Methods: Three computational models representing different body sizes were developed. An irrigated catheter ablation was simulated by delivering a 30 W pulse for 30 s to the endocardial surface of the anterior wall. Lesion sizes were then compared between two configurations of the dispersive patch (DP): an anterior (concordant) position and a posterior (discordant) position. (3) Results: Lesion size was consistently and significantly greater with concordant DP positioning compared to discordant positioning. Moreover, the magnitude of this difference decreased significantly with increasing body size, ranging from 0.65 ± 0.08 mm in the 35 kg swine model to 0.51 ± 0.06 mm in the human model. (4) Conclusions: Body size has a modest influence on the effect of dispersive patch positioning on RF lesion size. The potential advantage of a concordant DP configuration may be more significant in individuals with smaller body volume.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2025-09-24
DOI
https://doi.org/10.3390/bioengineering12101017
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Body Size Modulates the Impact of the Dispersive Patch Position During Radiofrequency Cardiac Ablation

Enrique Berjano, Ramiro M. Irastorza
Bioengineering
Cardiac Arrhythmias and Treatments
article

Body Size Modulates the Impact of the Dispersive Patch Position During Radiofrequency Cardiac Ablation

Enrique Berjano, Ramiro M. Irastorza
article en

Abstract

(1) Background: In the context of cardiac radiofrequency (RF) ablation, it has been proposed that positioning the dispersive patch (DP) concordantly with the orientation of the ablation electrode may enhance lesion size. The objective of this study is to investigate how individual body size may modulate the extent of this effect. (2) Methods: Three computational models representing different body sizes were developed. An irrigated catheter ablation was simulated by delivering a 30 W pulse for 30 s to the endocardial surface of the anterior wall. Lesion sizes were then compared between two configurations of the dispersive patch (DP): an anterior (concordant) position and a posterior (discordant) position. (3) Results: Lesion size was consistently and significantly greater with concordant DP positioning compared to discordant positioning. Moreover, the magnitude of this difference decreased significantly with increasing body size, ranging from 0.65 ± 0.08 mm in the 35 kg swine model to 0.51 ± 0.06 mm in the human model. (4) Conclusions: Body size has a modest influence on the effect of dispersive patch positioning on RF lesion size. The potential advantage of a concordant DP configuration may be more significant in individuals with smaller body volume.

BioengineeringVol. 12(10)
National Technological University (AR), Instituto de Fisica de Liquidos y Sistemas Biologicos (AR), Universitat Politècnica de València (ES), Universidad Nacional de La Plata (AR)
Universidad Tecnológica Nacional, Ministerio de Ciencia, Innovación y Universidades, Agencia Nacional de Promoción de la Investigación, el Desarrollo Tecnológico y la Innovación, European Regional Development Fund, Agencia Estatal de Investigación
Openalex Percentile: Top 60%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Body Size Modulates the Impact of the Dispersive Patch Position During Radiofrequency Cardiac Ablation — Enrique Berjano, Ramiro M. Irastorza · Bioengineering (2025) | TGRS Research Map | TGRS