Laboratory Validation of a Fully Automated Point-of-Care Device for High-Order Multiplexing Real-Time PCR Detection of Respiratory Pathogens

<b>Background/Objectives</b>: We have previously reported the engineering of a point-of-care (POC) system that fully automates the procedures for nucleic acid extraction and multiplexed real-time RT-PCR, with a major advantage of high-level multiplexing. In this study, we applied and validated the system in a respiratory tract infection setting. <b>Methods</b>: An automatic nested real-time RT-PCR assay was developed (POCm). It was a 40-plex assay that simultaneously detected 39 epidemiologically important respiratory pathogens in 1.5 h in the POC system. The analytical and clinical performance was evaluated. <b>Results</b>: The analytical sensitivities of the POCm assay were comparable to those of its single-plex counterparts performed manually on a bench-top. The minimum detectable concentrations ranged from 53 copies/mL to 5.3 × 10<sup>3</sup> copies/mL for all pathogen targets except hCoV-NL63 (5.3 × 10<sup>4</sup> copies/mL). The quantitative performance was demonstrated by the linear correlations between Ct values and input concentrations for all pathogen targets, with 24 of them demonstrating coefficients of correlation (r) greater than 0.9. The POCm assay was subsequently evaluated in 283 clinical samples. A high level of agreement (98.2-100%) was achieved for pathogen detection results between POCm and standard diagnostic methods. The POCm result was also fully concordant with the result of another commercial POC multiplex platform. For positive clinical samples, pairwise Ct values measured by POCm closely correlated with those of the bench-top reference method (r = 0.70). The feasibility of mutation genotyping of the viral subtype was further demonstrated. <b>Conclusions</b>: This study demonstrated the practicality of POCm for routine testing in clinical laboratories. Further clinical trials are being conducted to evaluate the clinical performance of the system.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2025-09-25
DOI
https://doi.org/10.3390/diagnostics15192445
Citations
1
Primary Topic
Respiratory viral infections research
Type
article
Field-Weighted Citation Impact
1.35

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Laboratory Validation of a Fully Automated Point-of-Care Device for High-Order Multiplexing Real-Time PCR Detection of Respiratory Pathogens

Parker Y. L. Tsang, Lok Ting Lau, Firaol Tamiru Kebede, Anita Yee et al.
1 citations
Diagnostics
Respiratory viral infections research
1.35
article

Laboratory Validation of a Fully Automated Point-of-Care Device for High-Order Multiplexing Real-Time PCR Detection of Respiratory Pathogens

Parker Y. L. Tsang, Lok Ting Lau, Firaol Tamiru Kebede, Anita Yee, Kelvin Kai‐Wang To, Deborah M. S. Tai, Teresa Chung, Johnson Y.N. Lau, Pete Y. M. Leung, Libby C. W. Li, Cyril Chik‐Yan Yip, Sunny L. H. Chu, Jonathan Hon-Kwan Chen, Carol Sze Ki Leung, Nancy B. Y. Tsui, Kwok-Yung Yuen, Manson Fok, Rosana Wing‐Shan Poon, Brenda Leung, Winston Wong
article en
1 citations

Abstract

<b>Background/Objectives</b>: We have previously reported the engineering of a point-of-care (POC) system that fully automates the procedures for nucleic acid extraction and multiplexed real-time RT-PCR, with a major advantage of high-level multiplexing. In this study, we applied and validated the system in a respiratory tract infection setting. <b>Methods</b>: An automatic nested real-time RT-PCR assay was developed (POCm). It was a 40-plex assay that simultaneously detected 39 epidemiologically important respiratory pathogens in 1.5 h in the POC system. The analytical and clinical performance was evaluated. <b>Results</b>: The analytical sensitivities of the POCm assay were comparable to those of its single-plex counterparts performed manually on a bench-top. The minimum detectable concentrations ranged from 53 copies/mL to 5.3 × 10<sup>3</sup> copies/mL for all pathogen targets except hCoV-NL63 (5.3 × 10<sup>4</sup> copies/mL). The quantitative performance was demonstrated by the linear correlations between Ct values and input concentrations for all pathogen targets, with 24 of them demonstrating coefficients of correlation (r) greater than 0.9. The POCm assay was subsequently evaluated in 283 clinical samples. A high level of agreement (98.2-100%) was achieved for pathogen detection results between POCm and standard diagnostic methods. The POCm result was also fully concordant with the result of another commercial POC multiplex platform. For positive clinical samples, pairwise Ct values measured by POCm closely correlated with those of the bench-top reference method (r = 0.70). The feasibility of mutation genotyping of the viral subtype was further demonstrated. <b>Conclusions</b>: This study demonstrated the practicality of POCm for routine testing in clinical laboratories. Further clinical trials are being conducted to evaluate the clinical performance of the system.

DiagnosticsVol. 15(19)
Macau University of Science and Technology (MO), Hong Kong Baptist University (HK), Hong Kong Polytechnic University (HK), University of Hong Kong (HK)
Hong Kong Baptist University
Openalex Percentile: Top 15%
Respiratory viral infections research
1.35
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.