Comparison of Aerosol Generation Between Bag Valve and Chest Compression-Synchronized Ventilation During Simulated Cardiopulmonary Resuscitation

<b>Background</b>: Cardiopulmonary resuscitation can generate aerosols, potentially exposing healthcare workers (HCWs) to infection. Bag valve ventilation (BV) is widely used but is prone to aerosol dispersion, whereas chest compression-synchronized ventilation (CCSV) maintains a closed respiratory circuit. In this study, we compared aerosol generation between CCSV and BV during chest compressions following endotracheal intubation in a simulated resuscitation setting. <b>Methods</b>: In a randomized crossover design, 12 sessions each of CCSV and BV were conducted on an intubated manikin undergoing mechanical chest compressions for 10 min. Aerosols with ≤5-μm diameter were generated using a saline nebulizer and measured every minute with a particle counter positioned 50 cm from the chest compression site. Bayesian linear regression of minute-by-minute log-transformed aerosol particle counts was used to estimate group differences, yielding posterior means, 95% credible intervals, and posterior probabilities. <b>Results</b>: The aerosol particle counts increased during the initial 3 min with the use of both methods. Thereafter, the aerosol particle counts with CCSV stabilized, whereas those with BV continued to increase. From 4 to 10 min, the posterior probability that CCSV generated fewer particles exceeded 0.98, peaking at 9 min. Both peak and time-averaged log-transformed aerosol particle counts were significantly lower with CCSV than with BV (<i>p</i> = 0.010 and <i>p</i> = 0.020, respectively). <b>Conclusions</b>: In this simulation, CCSV generated significantly fewer aerosols than BV did during chest compressions, with differences emerging after 4 min and persisting thereafter. Thus, CCSV may reduce aerosol exposure of HCWs, supporting its early implementation during resuscitation in infectious disease settings.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2025-09-25
DOI
https://doi.org/10.3390/jcm14196790
Primary Topic
Infection Control and Ventilation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of Aerosol Generation Between Bag Valve and Chest Compression-Synchronized Ventilation During Simulated Cardiopulmonary Resuscitation

Daun Choi, Young Taeck Oh, Hang A Park, Choung Ah Lee
Journal of Clinical Medicine
Infection Control and Ventilation
article

Comparison of Aerosol Generation Between Bag Valve and Chest Compression-Synchronized Ventilation During Simulated Cardiopulmonary Resuscitation

Daun Choi, Young Taeck Oh, Hang A Park, Choung Ah Lee
article en

Abstract

<b>Background</b>: Cardiopulmonary resuscitation can generate aerosols, potentially exposing healthcare workers (HCWs) to infection. Bag valve ventilation (BV) is widely used but is prone to aerosol dispersion, whereas chest compression-synchronized ventilation (CCSV) maintains a closed respiratory circuit. In this study, we compared aerosol generation between CCSV and BV during chest compressions following endotracheal intubation in a simulated resuscitation setting. <b>Methods</b>: In a randomized crossover design, 12 sessions each of CCSV and BV were conducted on an intubated manikin undergoing mechanical chest compressions for 10 min. Aerosols with ≤5-μm diameter were generated using a saline nebulizer and measured every minute with a particle counter positioned 50 cm from the chest compression site. Bayesian linear regression of minute-by-minute log-transformed aerosol particle counts was used to estimate group differences, yielding posterior means, 95% credible intervals, and posterior probabilities. <b>Results</b>: The aerosol particle counts increased during the initial 3 min with the use of both methods. Thereafter, the aerosol particle counts with CCSV stabilized, whereas those with BV continued to increase. From 4 to 10 min, the posterior probability that CCSV generated fewer particles exceeded 0.98, peaking at 9 min. Both peak and time-averaged log-transformed aerosol particle counts were significantly lower with CCSV than with BV (<i>p</i> = 0.010 and <i>p</i> = 0.020, respectively). <b>Conclusions</b>: In this simulation, CCSV generated significantly fewer aerosols than BV did during chest compressions, with differences emerging after 4 min and persisting thereafter. Thus, CCSV may reduce aerosol exposure of HCWs, supporting its early implementation during resuscitation in infectious disease settings.

Journal of Clinical MedicineVol. 14(19)
Hallym University Dongtan Sacred Heart Hospital (KR)
Openalex Percentile: Top 62%
Infection Control and Ventilation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Comparison of Aerosol Generation Between Bag Valve and Chest Compression-Synchronized Ventilation During Simulated Cardiopulmonary Resuscitation — Daun Choi, Young Taeck Oh, et al. · Journal of Clinical Medicine (2025) | TGRS Research Map | TGRS