CHARACTERIZATION OF A TEMPERATURE CONTROLLED ENCLOSURE AND UNCERTAINTY ANALYSIS

Temperature-controlled enclosures such as drying ovens, incubators, and sterilizers are widely used in laboratories and industrial settings. In such systems, it is essential that the temperature distribution within the working volume be as uniform as possible to ensure consistent exposure of samples to thermal conditions. This study presents a detailed methodology for the characterization of temperature-controlled volumes, based on metrological principles and international guidelines. A practical application on a drying oven is included, where temperature measurements were carried out using 15 calibrated type K thermocouples connected to a multi-channel data logger. The placement of sensors, the measurement procedure, and the evaluation of results including temporal fluctuation and spatial distribution are explained. Additionally, the uncertainty components affecting the measurement results were identified, and a complete uncertainty budget was calculated. The results show a systematic deviation from the setpoint, which is discussed in terms of the chamber's internal control behavior. This study aims to support calibration laboratories by providing a reference for both characterization practices and uncertainty evaluation.

Authors

Publication Details

Journal
Eskişehir Technical University Journal of Science and Technology A - Applied Sciences and Engineering
Published
2025-09-25
DOI
https://doi.org/10.18038/estubtda.1611528
Primary Topic
Scientific Measurement and Uncertainty Evaluation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

CHARACTERIZATION OF A TEMPERATURE CONTROLLED ENCLOSURE AND UNCERTAINTY ANALYSIS

Öznur Avcı, Can Ekici
Eskişehir Technical University Journal of Science and Technology A - Applied Sciences and Engineering
Scientific Measurement and Uncertainty Evaluation
article

CHARACTERIZATION OF A TEMPERATURE CONTROLLED ENCLOSURE AND UNCERTAINTY ANALYSIS

Öznur Avcı, Can Ekici
article en

Abstract

Temperature-controlled enclosures such as drying ovens, incubators, and sterilizers are widely used in laboratories and industrial settings. In such systems, it is essential that the temperature distribution within the working volume be as uniform as possible to ensure consistent exposure of samples to thermal conditions. This study presents a detailed methodology for the characterization of temperature-controlled volumes, based on metrological principles and international guidelines. A practical application on a drying oven is included, where temperature measurements were carried out using 15 calibrated type K thermocouples connected to a multi-channel data logger. The placement of sensors, the measurement procedure, and the evaluation of results including temporal fluctuation and spatial distribution are explained. Additionally, the uncertainty components affecting the measurement results were identified, and a complete uncertainty budget was calculated. The results show a systematic deviation from the setpoint, which is discussed in terms of the chamber's internal control behavior. This study aims to support calibration laboratories by providing a reference for both characterization practices and uncertainty evaluation.

Eskişehir Technical University Journal of Science and Technology A - Applied Sciences and EngineeringVol. 26(3)
Openalex Percentile: Top 74%
Scientific Measurement and Uncertainty Evaluation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.