Force in robotic thoracic surgery –a one year analysis of DaVinci 5 force feedback

Lack of tactile sensation has been a limitation of robotic-assisted thoracic surgery (RATS). The da Vinci 5 System (launched in 2024) integrates force feedback (FFB), a technology that measures instrument-tip forces and relays them to console hand controllers. This study characterizes forces applied during RATS and evaluates the impact of FFB. Da Vinci 5 system data for all RATS procedures using FFB instruments performed in the US between March 29, 2024 and April 30, 2025 were reviewed. Common thoracic surgeries were analyzed, including lung resections, mediastinal, esophageal, and diaphragmatic procedures. Mean instrument-tip forces (Newtons, N) were compared by procedure and instrument type. Forces and time spent above > 6.5 N were compared across FFB settings. Data from 444 procedures by 73 unique surgeons were analyzed. Median forces during lung resections (1.45 N anatomic resection, 1.42 N wedge) were significantly lower compared to mediastinal procedures (1.61 N), esophageal (1.74 N), and diaphragm surgery (1.59 N) (p < 0.001). Higher forces were measured with the use of retraction (Cadiere or Fenestrated Bipolar) than dissecting (Maryland) instruments (p < 0.001). Median forces decreased progressively with higher FFB settings (off, 1.79 N vs. low, 1.67 N vs. medium, 1.45 N vs. high, 1.32 N, p < 0.001). Proportion of duration at force > 6.5 N similarly declined with increased FFB setting (p < 0.001). FFB technology is associated with reduced average and peak instrument forces during RATS, particularly at medium and high settings. Retraction instruments experienced the highest forces. Further research is needed to define optimal force thresholds and clinical impact.

Authors

Institutions

Publication Details

Journal
Journal of Robotic Surgery
Published
2025-09-25
DOI
https://doi.org/10.1007/s11701-025-02781-9
Citations
1
Primary Topic
Surgical Simulation and Training
Type
article
Field-Weighted Citation Impact
2.01
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Force in robotic thoracic surgery –a one year analysis of DaVinci 5 force feedback

Robert E. Merritt, Brian Mitzman, Elliot L. Servais, Nestor Villamizar et al.
1 citations
Journal of Robotic Surgery
Surgical Simulation and Training
2.01
article

Force in robotic thoracic surgery –a one year analysis of DaVinci 5 force feedback

Robert E. Merritt, Brian Mitzman, Elliot L. Servais, Nestor Villamizar, Luis F. Tapias, John F. Lazar, Desmond M. D’Souza, Gretchen Purcell Jackson, Daniel Oh, Robert Mostellar, Peter J. Kneuertz
article en
1 citations

Abstract

Lack of tactile sensation has been a limitation of robotic-assisted thoracic surgery (RATS). The da Vinci 5 System (launched in 2024) integrates force feedback (FFB), a technology that measures instrument-tip forces and relays them to console hand controllers. This study characterizes forces applied during RATS and evaluates the impact of FFB. Da Vinci 5 system data for all RATS procedures using FFB instruments performed in the US between March 29, 2024 and April 30, 2025 were reviewed. Common thoracic surgeries were analyzed, including lung resections, mediastinal, esophageal, and diaphragmatic procedures. Mean instrument-tip forces (Newtons, N) were compared by procedure and instrument type. Forces and time spent above > 6.5 N were compared across FFB settings. Data from 444 procedures by 73 unique surgeons were analyzed. Median forces during lung resections (1.45 N anatomic resection, 1.42 N wedge) were significantly lower compared to mediastinal procedures (1.61 N), esophageal (1.74 N), and diaphragm surgery (1.59 N) (p < 0.001). Higher forces were measured with the use of retraction (Cadiere or Fenestrated Bipolar) than dissecting (Maryland) instruments (p < 0.001). Median forces decreased progressively with higher FFB settings (off, 1.79 N vs. low, 1.67 N vs. medium, 1.45 N vs. high, 1.32 N, p < 0.001). Proportion of duration at force > 6.5 N similarly declined with increased FFB setting (p < 0.001). FFB technology is associated with reduced average and peak instrument forces during RATS, particularly at medium and high settings. Retraction instruments experienced the highest forces. Further research is needed to define optimal force thresholds and clinical impact.

Journal of Robotic SurgeryVol. 19(1)
Saint Thomas West Hospital (US), Lahey Hospital and Medical Center (US), Saint Thomas Health (US), University of Miami (US), University of Utah (US), The Ohio State University Wexner Medical Center (US), Lahey Medical Center (US), City of Hope (US), Mayo Clinic in Arizona (US), Intuitive Surgical (United States) (US), Vanderbilt University Medical Center (US)
Openalex Percentile: Top 11%
Surgical Simulation and Training
2.01
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.