A Study on RF-Induced Heating of Passive Implantable Medical Devices at 5T MRI
This paper investigates the RF-induced heating of passive implantable medical devices (PIMDs) within the newly developed 5T whole-body MR system. Numerical and experimental studies identified the worst-case device length inside the ASTM phantom, which was then positioned at a clinically relevant location within an anatomically accurate human body model to evaluate RF-induced heating. The results indicate that the worst-case construct for RF-induced heating can be determined, and MR conditionality can be achieved for medical implants at 5T MRI.
Authors
- Jianfeng Zheng (ORCID: https://orcid.org/0000-0001-8998-9561)
- Mir Khadiza Akter
- Md Zahidul Islam (ORCID: https://orcid.org/0009-0003-0615-3665)
- Ao Shen
- Ji Chen (ORCID: https://orcid.org/0000-0003-4401-4714)
Institutions
- University of Houston (US)
Publication Details
- Published
- 2025-08-18
- DOI
- https://doi.org/10.1109/emcsipi52291.2025.11169859
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00