Comparison of Performances of Two Swept-Source Optical Coherence Tomography-based Biometers for Measurements of Eyes with Long Axial Lengths

Purpose: To compare the performances of two swept-source optical coherence tomography-based biometers for measuring ocular biometry and predicting postoperative refractive errors in eyes with long axial lengths (ALs).Methods: The ALs, anterior chamber depths (ACDs), corneal curvatures, and lens thicknesses of 23 eyes of 17 patients with ALs of ≥ 27 mm measured using the IOL Master<sup>®</sup> 700 and ARGOS<sup>®</sup> before cataract surgery were compared. The predicted and actual postoperative refractive errors were compared to assess the accuracy of the two devices.Results: The study participants comprised nine males (12 eyes) and eight females (11 eyes) with a mean age of 63.4 years. The two devices showed good agreement for ocular biometry, with high intraclass correlation coefficients for AL (0.994), ACD (0.972), corneal curvature (0.994), and lens thickness (0.943). The AL measured using the IOL Master<sup>®</sup> 700 was significantly higher than that measured using ARGOS<sup>®</sup> (<i>p</i> = 0.013). In contrast, the ACD measured using the IOL Master<sup>®</sup> 700 was lower than that measured using ARGOS<sup>®</sup> (<i>p</i> < 0.001). The mean prediction error for the IOL Master<sup>®</sup> 700 was significantly lower than that for the ARGOS<sup>®</sup> (<i>p</i> = 0.043). However, the mean absolute prediction errors did not differ significantly for the two devices (<i>p</i> = 0.601).Conclusions: The IOL Master<sup>®</sup> 700 and ARGOS<sup>®</sup> showed good agreement for eyes with long AL. However, significant differences in AL and ACD were observed. The two devices were comparable in predicting postoperative refractive errors.

Authors

Publication Details

Journal
Annals of Optometry and Contact Lens
Published
2025-09-25
DOI
https://doi.org/10.52725/aocl.2025.24.3.127
Citations
1
Primary Topic
Optical Coherence Tomography Applications
Type
article
Field-Weighted Citation Impact
0.67
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of Performances of Two Swept-Source Optical Coherence Tomography-based Biometers for Measurements of Eyes with Long Axial Lengths

Sang Beom Han, Ji Su Seong
1 citations
Annals of Optometry and Contact Lens
Optical Coherence Tomography Applications
0.67
article

Comparison of Performances of Two Swept-Source Optical Coherence Tomography-based Biometers for Measurements of Eyes with Long Axial Lengths

Sang Beom Han, Ji Su Seong
article en
1 citations

Abstract

Purpose: To compare the performances of two swept-source optical coherence tomography-based biometers for measuring ocular biometry and predicting postoperative refractive errors in eyes with long axial lengths (ALs).Methods: The ALs, anterior chamber depths (ACDs), corneal curvatures, and lens thicknesses of 23 eyes of 17 patients with ALs of ≥ 27 mm measured using the IOL Master<sup>®</sup> 700 and ARGOS<sup>®</sup> before cataract surgery were compared. The predicted and actual postoperative refractive errors were compared to assess the accuracy of the two devices.Results: The study participants comprised nine males (12 eyes) and eight females (11 eyes) with a mean age of 63.4 years. The two devices showed good agreement for ocular biometry, with high intraclass correlation coefficients for AL (0.994), ACD (0.972), corneal curvature (0.994), and lens thickness (0.943). The AL measured using the IOL Master<sup>®</sup> 700 was significantly higher than that measured using ARGOS<sup>®</sup> (<i>p</i> = 0.013). In contrast, the ACD measured using the IOL Master<sup>®</sup> 700 was lower than that measured using ARGOS<sup>®</sup> (<i>p</i> < 0.001). The mean prediction error for the IOL Master<sup>®</sup> 700 was significantly lower than that for the ARGOS<sup>®</sup> (<i>p</i> = 0.043). However, the mean absolute prediction errors did not differ significantly for the two devices (<i>p</i> = 0.601).Conclusions: The IOL Master<sup>®</sup> 700 and ARGOS<sup>®</sup> showed good agreement for eyes with long AL. However, significant differences in AL and ACD were observed. The two devices were comparable in predicting postoperative refractive errors.

Annals of Optometry and Contact LensVol. 24(3)
Openalex Percentile: Top 26%
Optical Coherence Tomography Applications
0.67
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.