Measurement accuracy of CT systems: The importance of calibration phantoms

This study aims to evaluate the measurement accuracy of computed tomography (CT) systems, focusing on the necessity of using calibration phantoms for enhanced precision. Both clinical CT and micro-CT systems were evaluated using a specially designed two-ball phantom, which provides a reliable reference for spatial resolution and geometric accuracy. The study involved scanning the phantom with two micro-CT devices (the oversize micro-CT SkyScan 1173 and the high-resolution micro-CT SkyScan 1272) and a clinical CT device, a third-generation dual-source CT scanner (SOMATOM Force), measuring the distance between the centres of two ruby balls. The results showed significant differences in measurement accuracy between the devices. The high-resolution micro-CT provided the most consistent measurements with minimal variance, indicating its superiority in applications requiring high precision. In contrast, the oversize micro-CT exhibited larger errors, particularly at smaller voxel sizes, suggesting that internal calibration affected its accuracy. The dual source CT system had the smallest mean error but a larger standard deviation, indicating less consistency compared to micro-CT systems. Calibration with the two-ball phantom improved measurement accuracy across all devices. This improvement underscores the importance of using calibration phantoms to ensure accurate measurements, especially in fields that require high precision, such as clinical diagnostics and materials science. We concluded that routine calibration with phantoms is essential to achieve high measurement accuracy in CT imaging, thereby increasing the reliability of CT-based analyses in various disciplines.

Authors

Institutions

Publication Details

Journal
PLoS ONE
Published
2025-09-25
DOI
https://doi.org/10.1371/journal.pone.0332263
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Measurement accuracy of CT systems: The importance of calibration phantoms

Anton G. Windfelder, Jan Scherberich, Gabriele Anja Krombach, Jessica Steinbart
PLoS ONE
Advanced X-ray and CT Imaging
article

Measurement accuracy of CT systems: The importance of calibration phantoms

Anton G. Windfelder, Jan Scherberich, Gabriele Anja Krombach, Jessica Steinbart
article en

Abstract

This study aims to evaluate the measurement accuracy of computed tomography (CT) systems, focusing on the necessity of using calibration phantoms for enhanced precision. Both clinical CT and micro-CT systems were evaluated using a specially designed two-ball phantom, which provides a reliable reference for spatial resolution and geometric accuracy. The study involved scanning the phantom with two micro-CT devices (the oversize micro-CT SkyScan 1173 and the high-resolution micro-CT SkyScan 1272) and a clinical CT device, a third-generation dual-source CT scanner (SOMATOM Force), measuring the distance between the centres of two ruby balls. The results showed significant differences in measurement accuracy between the devices. The high-resolution micro-CT provided the most consistent measurements with minimal variance, indicating its superiority in applications requiring high precision. In contrast, the oversize micro-CT exhibited larger errors, particularly at smaller voxel sizes, suggesting that internal calibration affected its accuracy. The dual source CT system had the smallest mean error but a larger standard deviation, indicating less consistency compared to micro-CT systems. Calibration with the two-ball phantom improved measurement accuracy across all devices. This improvement underscores the importance of using calibration phantoms to ensure accurate measurements, especially in fields that require high precision, such as clinical diagnostics and materials science. We concluded that routine calibration with phantoms is essential to achieve high measurement accuracy in CT imaging, thereby increasing the reliability of CT-based analyses in various disciplines.

PLoS ONEVol. 20(9)
Biruni University (TR), Fraunhofer Institute for Molecular Biology and Applied Ecology (DE)
Openalex Percentile: Top 75%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Measurement accuracy of CT systems: The importance of calibration phantoms — Anton G. Windfelder, Jan Scherberich, et al. · PLoS ONE (2025) | TGRS Research Map | TGRS