Facial video photoplethysmography for measuring average and quasi-instantaneous heart rate: a pilot validation study
In predominantly young, white, seated subjects, the tested vPPG technology provided highly accurate HR measurements, both when the values were averaged over 60 s and in the case of short-term values averaged over 10 s or quasi-instantaneous values averaged over 4 s. To our knowledge, this is the first study on vPPG technology to examine quasi-instantaneous HR measurements (averaged over periods shorter than 5 s). The results should be confirmed in a larger study with greater diversity in age, skin tone, and lighting conditions.
Authors
- Bartłomiej Paleczny (ORCID: https://orcid.org/0000-0001-9180-9259)
- Tymoteusz Okupnik (ORCID: https://orcid.org/0000-0003-3194-9190)
- Beata Ponikowska (ORCID: https://orcid.org/0000-0002-6044-5322)
- Leszek Pstras (ORCID: https://orcid.org/0000-0001-5705-6142)
Institutions
- Wroclaw Medical University (PL)
- Institute of Biocybernetics and Biomedical Engineering (PL)
- Polish Academy of Sciences (PL)
Publication Details
- Journal
- Frontiers in Physiology
- Published
- 2025-09-24
- DOI
- https://doi.org/10.3389/fphys.2025.1638809
- Citations
- 1
- Primary Topic
- Non-Invasive Vital Sign Monitoring
- Type
- article
- Field-Weighted Citation Impact
- 0.67