Automated computation of detectability index and generation of contrast–detail curves for CT protocol optimization

<i>Objective.</i>The aim of this study was to develop an automatic method for generating a detectability index (<i>d'</i>)-based contrast-detail (<i>C</i>-<i>D</i>) curve across multiple object sizes and contrasts, and to evaluate its performance under varying tube current settings and reconstruction filter types.<i>Approach.</i>To compute<i>d'</i>for a given object size and contrast, the task-transfer function and noise power spectrum were obtained from ACR 464 computed tomography (CT) phantom images acquired at tube currents of 80, 120, 160 and 200 mA, using Edge, Lung, and Soft filter types. The task objects were varied in size (1-15 mm) and contrast levels (1-15 HU) with both flat and Gaussian signal types. For each defined task object,<i>d'</i>was calculated using a non-prewhitening model observer. This process was iterated for every predefined task function across multiple object sizes and contrasts, resulting in a<i>d'</i>map corresponding to the synthetic low-contrast images. A<i>C</i>-<i>D</i>curve was then generated using a<i>d'</i>cut-off value defined by the user. For comparison, a separate<i>C</i>-<i>D</i>curve was generated based on visual assessment by five human observers (HOs).<i>Main results.</i>The automated method successfully computed<i>d'</i>values and arranged synthetic low-contrast images into a grid according to object size and contrast.<i>C</i>-<i>D</i>curves using<i>d'</i>cut-off values of 3 or 4 most closely reflected HOs performance. For tube current variations, increasing the current led to higher detectability. For filter type variations, the Lung filter resulted in relatively lower detectability compared to the Edge and Soft filters.<i>Significance</i>. An automated method to calculate<i>d'</i>across a wide range of object sizes and contrasts, and to generate a<i>d'</i>-based<i>C</i>-<i>D</i>curve for CT protocol optimization was developed. The results were consistent with HO trends and effectively captured detectability changes across different imaging parameters.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-09-23
DOI
https://doi.org/10.1088/1361-6560/ae0ab0
Citations
2
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
1.35
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Automated computation of detectability index and generation of contrast–detail curves for CT protocol optimization

Heri Sutanto, Ariij Naufal, Chai Hong Yeong, Choirul Anam et al.
2 citations
Physics in Medicine and Biology
Advanced X-ray and CT Imaging
1.35
article

Automated computation of detectability index and generation of contrast–detail curves for CT protocol optimization

Heri Sutanto, Ariij Naufal, Chai Hong Yeong, Choirul Anam, Geoff Dougherty, Kusworo Adi
article en
2 citations

Abstract

<i>Objective.</i>The aim of this study was to develop an automatic method for generating a detectability index (<i>d'</i>)-based contrast-detail (<i>C</i>-<i>D</i>) curve across multiple object sizes and contrasts, and to evaluate its performance under varying tube current settings and reconstruction filter types.<i>Approach.</i>To compute<i>d'</i>for a given object size and contrast, the task-transfer function and noise power spectrum were obtained from ACR 464 computed tomography (CT) phantom images acquired at tube currents of 80, 120, 160 and 200 mA, using Edge, Lung, and Soft filter types. The task objects were varied in size (1-15 mm) and contrast levels (1-15 HU) with both flat and Gaussian signal types. For each defined task object,<i>d'</i>was calculated using a non-prewhitening model observer. This process was iterated for every predefined task function across multiple object sizes and contrasts, resulting in a<i>d'</i>map corresponding to the synthetic low-contrast images. A<i>C</i>-<i>D</i>curve was then generated using a<i>d'</i>cut-off value defined by the user. For comparison, a separate<i>C</i>-<i>D</i>curve was generated based on visual assessment by five human observers (HOs).<i>Main results.</i>The automated method successfully computed<i>d'</i>values and arranged synthetic low-contrast images into a grid according to object size and contrast.<i>C</i>-<i>D</i>curves using<i>d'</i>cut-off values of 3 or 4 most closely reflected HOs performance. For tube current variations, increasing the current led to higher detectability. For filter type variations, the Lung filter resulted in relatively lower detectability compared to the Edge and Soft filters.<i>Significance</i>. An automated method to calculate<i>d'</i>across a wide range of object sizes and contrasts, and to generate a<i>d'</i>-based<i>C</i>-<i>D</i>curve for CT protocol optimization was developed. The results were consistent with HO trends and effectively captured detectability changes across different imaging parameters.

Physics in Medicine and BiologyVol. 70(19)
Diponegoro University (ID), California State University, Channel Islands (US), Taylor's University (MY)
Openalex Percentile: Top 17%
Advanced X-ray and CT Imaging
1.35
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.