Balloon Shaft–Assisted Guide Catheter Exchange Without PTCA Wire Removal
We describe a balloon shaft-assisted technique for guide catheter exchange without removing the in-situ PTCA wire. In a complex LCX in-stent restenosis case with poor guide support, a 6 Fr EBU catheter was successfully advanced over the metallic shaft of a monorail balloon tracked on a 0.014″ PTCA wire. Bench testing confirmed the shaft's ability to mimic a stiff 0.035″ wire, providing adequate support for safe catheter exchange. This technique eliminates the need for anchoring or specialized wires and offers a practical, resource-efficient solution for guide exchange, particularly in settings with limited device availability.
Authors
- Desabandhu Vinayakumar (ORCID: https://orcid.org/0000-0002-1956-8937)
- Kapil Rajendran (ORCID: https://orcid.org/0000-0003-0247-064X)
Institutions
- T. D. Medical College (IN)
- Software Training and Development Centre (IN)
Publication Details
- Journal
- Catheterization and Cardiovascular Interventions
- Published
- 2025-09-23
- DOI
- https://doi.org/10.1002/ccd.70216
- Primary Topic
- Central Venous Catheters and Hemodialysis
- Type
- article
- Field-Weighted Citation Impact
- 0.00