Comparative evaluation of two automated immunoassays for serum thyroglobulin quantification

Background: We evaluated the clinical performance of in vitro diagnostic devices for quantifying thyroglobulin (Tg), which is a key marker for monitoring and treating thyroid dysfunction. The recently launched Siemens Atellica IM Tg assay was compared with the established Roche Elecsys Tg II assay using residual serum samples from routine testing.Methods: The precision, linearity, limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ), and reference ranges were assessed using the Siemens Atellica IM Tg assay. In total, 681 patient serum samples were analyzed to compare the results with those of the Roche Elecsys Tg II assay for correlation and concordance evaluations across the clinical ranges.Results: The precision coefficients of variation, linearity, LoB, LoD, and LoQ for the Siemens Atellica IM Tg assay met the manufacturer’s specifications across all concentrations. All data points and confidence intervals were within the allowable deviation from linearity. The correlation between the assays was excellent (Pearson’s r=0.997). In the low range, concordance was 83%, whereas in the normal range, it reached 98%. In contrast, the highly abnormal range exhibited a concordance of 65%, resulting in an overall concordance rate of 88%. Weighted kappa values (κ=0.79–0.82) demonstrated moderate-to-strong agreement.Conclusions: The Siemens Atellica IM Tg assay showed performance consistent with the precision, linearity, LoB, LoD, LoQ, and reference ranges claimed by the manufacturer. It demonstrated a strong correlation and good overall concordance with Roche Elecsys Tg II. Lower concordance in the highly abnormal range suggests a potential limitation.

Authors

Institutions

Publication Details

Journal
Kosin Medical Journal
Published
2025-09-23
DOI
https://doi.org/10.7180/kmj.25.122
Primary Topic
Particle accelerators and beam dynamics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative evaluation of two automated immunoassays for serum thyroglobulin quantification

Kyoung Ho Roh, Hyunyong Hwang
Kosin Medical Journal
Particle accelerators and beam dynamics
article

Comparative evaluation of two automated immunoassays for serum thyroglobulin quantification

Kyoung Ho Roh, Hyunyong Hwang
article en

Abstract

Background: We evaluated the clinical performance of in vitro diagnostic devices for quantifying thyroglobulin (Tg), which is a key marker for monitoring and treating thyroid dysfunction. The recently launched Siemens Atellica IM Tg assay was compared with the established Roche Elecsys Tg II assay using residual serum samples from routine testing.Methods: The precision, linearity, limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ), and reference ranges were assessed using the Siemens Atellica IM Tg assay. In total, 681 patient serum samples were analyzed to compare the results with those of the Roche Elecsys Tg II assay for correlation and concordance evaluations across the clinical ranges.Results: The precision coefficients of variation, linearity, LoB, LoD, and LoQ for the Siemens Atellica IM Tg assay met the manufacturer’s specifications across all concentrations. All data points and confidence intervals were within the allowable deviation from linearity. The correlation between the assays was excellent (Pearson’s r=0.997). In the low range, concordance was 83%, whereas in the normal range, it reached 98%. In contrast, the highly abnormal range exhibited a concordance of 65%, resulting in an overall concordance rate of 88%. Weighted kappa values (κ=0.79–0.82) demonstrated moderate-to-strong agreement.Conclusions: The Siemens Atellica IM Tg assay showed performance consistent with the precision, linearity, LoB, LoD, LoQ, and reference ranges claimed by the manufacturer. It demonstrated a strong correlation and good overall concordance with Roche Elecsys Tg II. Lower concordance in the highly abnormal range suggests a potential limitation.

Kosin Medical JournalVol. 40(3)
National Health Insurance Service Ilsan Hospital (KR), National Health Insurance Service (KR)
Openalex Percentile: Top 66%
Particle accelerators and beam dynamics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Comparative evaluation of two automated immunoassays for serum thyroglobulin quantification — Kyoung Ho Roh, Hyunyong Hwang · Kosin Medical Journal (2025) | TGRS Research Map | TGRS