Methods and equipment for metrological certification of phase-measuring equipment
The paper analyzes the main components characterizing the accuracy of phase measurement methods: — basic error, at equal levels of the signals under study; — amplitude-phase error at unequal levels of the signals under study. The hardware for determining these error components is: — a developed microcontroller two-phase phase calibrator based on DDS generators, the accuracy of setting phase shifts of which in the frequency range of up to 1 MHz is hundredths of a degree; — a calibrated attenuator D2-31 with fixed attenuation values of 10 dB, providing a phase error at a frequency of 1 MHz of no more than 0,03º.
Authors
- L. R. Grigoryan
- R. L. Grigoryan
- N. M. Bogatov
Publication Details
- Journal
- Omsk Scientific Bulletin
- Published
- 2025-01-01
- DOI
- https://doi.org/10.25206/1813-8225-2025-195-111-117
- Primary Topic
- Advanced Measurement and Metrology Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00